13
If the Last Fitting Sensor does not light with four (4) strips of paper, use a 3mm Metric Hex
Wrench (T Handle preferred) to adjust the sensor position upwards (turn the adjustment screw
clockwise) until the Last Fitting Sensor light turns on. (Figure 9.2 & 9.3)
Figure 9.2 – Adjusting Sensor Position
Figure 9.3 – Adjustment Screw Detail
After verifying that the Last Fitting Sensor light is lit with four (4) pieces of paper, add the fifth
strip of paper between the magazine and the top of the magazine rack. (Figure 9.4)
The Last Fitting Sensor light should go out. If the light goes out, press firmly on the top of the
magazine as shown (Figure 9.5) to check if the light will come back on.
Figure 9.4 – Five strips of paper
Figure 9.5 – Pressing on magazine
Adjustment
Screw
Last Fitting
Sensor Light
Off
Five strips
of paper