Power on

Short press power on/off key to start the meter.  The whole 

screen displayed within 1 second. 
Power off

Long press the power on/off key to power off the meter. 

Auto power off in 10 minutes if without any operation. 
LCD backlight

Short press          key to control if turn on the backlight or not.

UNIT

MODE

MODE

Press           to switch unit (ppm, ppb),

Press           to switch from Dissolved Hydrogen

concentration mode to Reduction Potential mode

With the sensor soaking solution display: 0.000ppm. 

Function Setting

Battery

Power on/off and Backlight

Electrode using and maintenance

Use the coin to remove the battery cover by following the “-” direction.
Insert 2*7AAA batteries in the battery compartment with the correct 
Positive and Negative poles
Tight the battery cover by following the “+” direction.

Electrode Installation

Shell

The direction of the red dot marker on the Electrode and the salient 
point on the unit should be the same when install the electrode to the 
unit. Remove the electrode by pulling out the clamp on the electrode.

If  the meter accidentally touches water or other fluids, please wipe it 
clean immediately. The meter function is not affected usually. But if 
affected and cannot use (possibly because of the battery cover is not 
tight and penetrated by water), please contact our company or 
CLEAN local customer service center. They will be very glad to help you.

Notice

1.Please keep the metallic diaphragm always clean. Make sure the sensor 
protection bottle is filled up with soaking solution. Note: Other solutions or tap 
water can not be used as protection solution.
2. Please rinse off the sensor tip with clean water after each use and cover 
the protection bottle back. Please also check the volume of the soaking 
solution. If not enough, please be sure to add more.
Note: After each use, if there is some moisture or solution liquid on and near 
the metallic diaphragm, random numbers may get displayed. Rinse off the 
tip with clean water and place it back in sensor protecting cap. Make sure it 
displays”0” before shutting off power.
3. If there is smudginess and mineral elements adhere on the metallic 
diaphragm, the responsive time will be deduced and readings might drop. 
Please use the wiping paper to polish the electrode metallic diaphragm. 
Make sure the sensor tip is very clean. 
After cleaning, please must keep the sensor tip in soaking solution. If the 
reading is not 0.000ppm when in soaking solution, please change the sensor.
4. Life expectancy of an electrode depends on frequency of use. Generally, 
it should be replaced in 1 year.Generally, it should be replaced in 1 year. 

Cleaning order of the electrode metallic diaphragm:

1. Spin off the sensor protection bottle (please pay attention not to spill out the 
soaking solution).
2. Hold the wiping paper (with the rough surface up) and drop a few drops of 
water on the paper.
3. Horizontally contact the bottom of the metallic diaphragm with the wiping 
paper which having water on it. Rub rotating as you writing the Arabic 
numeral “8” for around 10 times.
4. Rinse off the tip with clean water and place it back in sensor protecting 
cap. Make sure it displays”0” before shutting off power.  

BNC

+

-

M

SET

Re M.

MR

DH200

+

-

UNIT

Soaking solution

(CS6080S)

Base line 

Sensor

protection

bottle 

Electrode

metallic

diaphragm 

CS6080

Soak the sensor in the bottle.

Fill up but don’t overflow. 

CW100 

wiping paper 

Polish the electrode 

metallic diaphragm 

Authorization must be obtained from CLEAN Instruments Customer Service 

Center to issue a  RIR number before returning items for any reason. When 

applying for authorization, please  nclude date requiring the reason of return. 

Instruments must be carefully packed to prevent  damage in shipment and 

insured against possible damage or loss. CLEAN Instruments will not  be 

responsible for any damage resulting from careless or insufficient packing.

Warning: Damage as a result of inadequate packaging is the User / 

distributor’s responsibility.

Please follow the guidelines below before transporting.

Use the original packaging material if possible, when transporting back the 

unit for repair. 

Otherwise wrap it with bubble pack and use a corrugated box for better 

protection. Include a brief description of any faults suspected for the 

convenience of Customer Service Center, if possible . If there are any 

questions, feel free to contact our Customer Service Center or distributors.

Warranty 

Return Of Malfunction Instruments 

Guidelines Or Returning Unit For Repair

CLEAN Instruments warrants this product to be free from significant deviations 

in material and  workmanship for a period of one year from the date of 

purchase. If repair is necessary and has not been the result of abuse or 

misuse within the warranty period, please return to CLEAN  Instruments and 

amendment will be made without any charge. CLEAN Instruments Customer 

Service Center will determine if product problem is due to deviations or 

customer abuse. Out of warranty products will be repaired on a charge basis.