X144D e-FlowMeter

Quick Start Programming 

SETTINGS – OUTPUT (Continued)

Set Pulse Weight?

NEXT

01.000

Default Pulse Weight is 1.00. Use arrow keys to increase or decrease Pulse Weight. 
Flow Rate = Number of Pulses x Pulse Weight / Time.  Select 

to continue.

NEXT

Set Flow Cutoff

NEXT

030.00

This setting is used to set a minimum flow reading that is displayed on screen before displaying

zero flow.  This is used when noise or vibration is causing erratic low flow readings when there is
no flow. Select 

to continue.

NEXT

Are you sure you

want to save

changes?

YES

NO

To save changes select 

To discard changes and advance to next menu select  

.

YES

NO

Do you want to test

the 4-20mA Loop?

YES

NO

To test 4 and 20mA outputs select 

. To return to settings menu select  

.

YES

NO

NEXT

0655

In this menu, a 4mA output signal is generated. To calibrate 4mA output to external device, use

arrow keys to increase or decrease 4mA output. A Digital Multimeter to measure 4mA output.
Select 

to continue to 20mA test.

NEXT

EXIT

3273

In this menu, a 20mA output signal is generated. To calibrate 20mA output to external device,

use arrow keys to increase or decrease 20mA output. A Digital Multimeter to measure 20mA out-
put Select 

to return to Settings Menu.

EXIT