Civco GUS G25-OER Operator'S Manual Download Page 7

GUS Vapor Control System™

 

 

 

 

M0015_F 

G25-OER 

 

 

 

 

 

 

Installation 

 

CAUTION:

   

It will require two people to lift the system and install it on the wall.

 

1.

 

The G25-OER is 26.5” wide and will require at least an additional one inch clearance on 

either side. There must be a minimum of 4” of clearance above the unit to allow for 

adequate air circulation. 

 

2.

 

Locate the center point of the OER-Pro on the wall behind the unit.  Mark the wall at 67” 

above the floor at the center point. 

 

3.

 

Using a level draw a horizontal line 13.25” on each side of the center mark. 

 

4.

 

Locate the bottom of the Z-bracket along the horizontal line.  Mark the location of the six 

holes on the Z-bracket against the wall.   

 

5.

 

If the wall is of wood or metal stud construction, locate studs and drill new holes through 

the Z-bracket to align with the center of the studs.   

 

6.

 

Alternatively

, mount a piece of ¾” plywood cut to 38” high x 26.5” wide.  Screw this directly 

into the studs.  The Z-bracket can then be screwed directly into the plywood. 

 

7.

 

In the event that studs cannot be located use the EZ Anchors that are supplied and screw 

into the wall at each mark.   

 

NOTE    

Mount the Z-bracket with open side upward

.

 

 

8.

 

Using the lifting straps lift the GUS system up and over the matching ‘Z’ bar on the wall. It is 

a tight fit and you must make sure the two Z-brackets are securely interlocked before 

releasing hold on the unit.  

 

9.

 

The unit has two “L” brackets. One (upper “L” bracket) is attached to the lower section of 

the blower housing.  The second (lower “L” bracket) is attached to the side panels of the 

G25-OER. Mark the location of the bracket holes for both upper and lower “L” brackets on 

the wall. Either use EZ anchors supplied or screw directly into plywood. If using EZ anchors 

see step 11. 

 

10.

 

If wall is plaster or sheet rock, REMOVE the “L” brackets from the unit by removing the 

mounting screws.  This will allow you to install the EZ anchors at marks made in step 10. 

 

11.

 

Install EZ anchors in the wall and re-attach “L” brackets to the unit with screws removed in 

step 10.   

 

12.

 

Using supplied screws attach unit to wall through “L” brackets into EZ anchors. 

 

13.

 

Plug into a 110-volt wall outlet. The unit is now ready for use.     

 

To review this or if you have any questions, please call CIVCO Medical Solutions at  

1-800-445-6741

 

Summary of Contents for GUS G25-OER

Page 1: ...dical Solutions All other trademarks are property of their respective owners Printed in the USA Model G25 OER Vapor Control System Operator s Manual CIVCO Medical Solutions 102 First Street South Kalo...

Page 2: ...ions 2 Ventilation Requirements 2 Supply Ratings 2 Warning Symbols 3 Signal Words 3 Quick Troubleshoot 4 Precautions 4 Installation and Setup 5 Upon Delivery 5 Installation 6 Operation 7 Maintenance 7...

Page 3: ...power before replacing filter or when servicing the unit Refer to the MSDS of the high level disinfectant that you are using for additional safety precautions Environmental Conditions For indoor use...

Page 4: ...ignal Words The following signal words are used throughout this manual WARNING Indicates a potentially hazardous situation which if not avoided could result in death or serious injury CAUTION Indicate...

Page 5: ...GUSG25 OER Vapor Control System check the following 1 The system is being left on at all times 2 When fresh disinfectant is being poured into the OER Pro that the G25 OER is running on high speed 3 F...

Page 6: ...maged or missing or if you have any questions call CIVCO Medical Solutions at 1 800 445 6741 Do not use the system Items shipped with your G25 OER for installation One Z bracket for wall mounting Twel...

Page 7: ...ated use the EZ Anchors that are supplied and screw into the wall at each mark NOTE Mount the Z bracket with open side upward 8 Using the lifting straps lift the GUS system up and over the matching Z...

Page 8: ...veryday use will have a six month life Call CIVCO Medical Solutions at 1 800 445 6741 to purchase additional filters CAUTION Disconnect the system s power supply before removing the filter 1 Turn the...

Page 9: ...n spots may be cleaned with a mild solvent if extra care is taken Disposing Neutralizing Use the two 2 gal containers supplied with your G25 OER Dispose half of the contents of a bottle of neutralizer...

Page 10: ...IS EXPRESSLY IN LIEU OF ALL OTHER WARRANTIES GUARANTEES OBLIGATIONS OR OTHER LIABILITIES EXPRESSED OR IMPLIED AND THERE ARE NO WARRANTIES WHICH EXTEND BEYOND THE DESCRIPTION ON THE FACE HEREOF WARRAN...

Page 11: ...losed 40 watts Current Running with door open 75 watts Filters Carbon filter Activated carbon Fan Blower Automatic two speed Construction Exhaust System Powder coated carbon steel Door Acrylic Baffle...