12. Without regular recalibration, the results may not be accurate. Every CTD505 is 

calibrated right after production. See “Warranty and Recalibration” for details.

13. Keep the CTD505 away from sunlight and dusty places.

Important Notice:
The CTD505 does not require a probe cover during use. It is important to 
clean and disinfect the CTD505 before and after each use. See “Cleaning and 
Disinfection” for details.
For measurement:
 •  When asleep, pressure on the ear may cause higher-than-normal temperature. 

Wait a few minutes, and then measure again.

 •  Clean your external ear before measuring. To ensure an accurate result, keep 

your ear free from obstructions or excess earwax.

 •  The measuring results from the left and right ears may differ slightly. We 

recommend measuring in the same ear.

 •  Before measuring, the user and the CTD505 should be in a steady room 

temperature for at least 30 minutes. The user should also rest for 30 minutes 
after exercising.

HOW TO  OPERATE

1. Press the 

|

/O

 button. The backlight is switched on for 

approximately 2 seconds. The LCD screen displays all digital 
segments.

2.  If the CTD505 has saved past results, the LCD screen 

automatically displays the latest result.

3. When 

 fl ashes with a short beep, the CTD505 is ready to use.

1.  To improve accuracy, follow the illustrations below. For adults, pull the ear up 

and back. For infants and children, pull the ear back.

Adults

Infants and Children

2.  Fit the probe tip into the ear canal, and then press the 

START

 button. Release the button and wait a few seconds 

until you hear a long beep. 

3.  Take the CTD505 out of the ear canal. The LCD displays 

the latest measuring result. The backlight is switched on for approximately 5 
seconds. Without any operation, the CTD505 automatically shuts off after 1 
minute.

Note: 

Before re-measurement, wait 1 – 3 minutes.

Summary of Contents for CTD 505

Page 1: ...INSTRUCTION MANUAL FOR DIGITAL EAR THERMOMETER Model CTD 505 Antibacterial Model English 1401 ...

Page 2: ...ture of people of any age from the auditory canal in a home setting 2 Do not use the CTD505 as a replacement for medical advice Consult your doctor if you have any concerns 3 ONLY the CTD505 probe tip is waterproof Do not immerse it in water or any other liquid 4 Do not drop the CTD505 and avoid any strong impact 5 Do not use the CTD505 on any body part other than your ear 6 Do not move or shake y...

Page 3: ...the same ear Before measuring the user and the CTD505 should be in a steady room temperature for at least 30 minutes The user should also rest for 30 minutes after exercising HOW TO OPERATE 1 Press the O button The backlight is switched on for approximately 2 seconds The LCD screen displays all digital segments 2 If the CTD505 has saved past results the LCD screen automatically displays the latest...

Page 4: ...splays the previous measuring result release the START button 3 To check memory erasing turn the CTD505 off and enter Memory Recall Mode again BATTERY REPLACEMENT flashes You can still measure but the accuracy may not impaired lights or no reaction at all Change the battery as soon as possible Note Change the battery away from small children and any source of heat 1 Press the upper part of the bat...

Page 5: ...prevent water getting into the CTD505 Use an alcohol swab to clean the probe area in a circular motion for 15 seconds around 20 times DISINFECTION Probe Exterior Area 1 Hold the CTD505 facing down to prevent water getting into the CTD505 Use an alcohol swab to clean the probe area in a circular motion for 15 seconds around 20 times 2 Then use the same alcohol swab to wipe the exterior area back an...

Page 6: ... lead to inaccurate results or malfunction Maintenance 1 Do not disassemble or change any parts of the CTD505 Such action may reduce its accuracy 2 Do not drop the CTD505 and avoid any strong impact ABOUT THE EAR THERMOMETER Normal body temperature differs by body parts time and age Here are two tables for your reference 1 Normal Body Temperature Range Age Site 0 2 years old 3 10 years old 11 65 y...

Page 7: ...measure Remove the worn out battery Insert a new CR2032 battery as soon as possible Without enough energy the CTD505 cannot function as it should lights Battery power is too low to measure Remove the worn out battery Insert a new CR2032 battery as soon as possible Too high ambient temperature Ambient temperature is higher than 40 0 C Leave the CTD505 in a room around 10 0 C 40 0 C for 30 minutes T...

Page 8: ... while measuring Repeat the measuring steps again If the error persists return the CTD505 to your local distributor OTHER INFORMATION This product conforms to the provisions of the EC directive 93 42 EEC Medical Device Directive And also complies with following standards included but not limited Safety standard EN 60601 1 Medical electrical equipment part 1 General requirements for safety EMC stan...

Page 9: ...gue The Netherlands Warranty Recalibration Warranty For One Year from Manufacturing Date CITIZEN SYSTEMS JAPAN Co Ltd grants a 1 year warranty on the product commencing on the date of purchase Within the limited time we will eliminate any defects in the appliance resulting from faults in the material or workmanship free of charge Please note that this service does not cover damage caused by misuse...

Page 10: ...39 52 mm Weight without battery 66 80 g Measurement range 34 0 C 43 0 C Low temperature display Temperature 34 0 C display L C High temperature display Temperature 43 0 C display H C Display resolution 0 1 C Lab accuracy 35 5 C 42 0 C 0 2 C Other range 0 3 C Clinical accuracy repeatability 0 3 C Memory sets 12 Auto power off 1 minute Battery CR2032 lithium battery 3V 1 Battery life 1000 times Oper...

Page 11: ...n such environments Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Conducted RF IEC 61000 4 6 Radiated RF IEC 61000 4 3 3 Vrms 150 MHz to 80 GHz 3 V m 80 MHz to 2 5 GHz Not Applicable 3 V m Portable and mobile RF communications equipment should be used no closer to any part of the device including cables than the recommended separation distance calculated ...

Page 12: ... 50 60 Hz magnetic field IEC 61000 4 8 3 A m 3 A m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment Recommended separation distances between portable and mobile RF communication equipment and the device The device is intended for use in an electromagnetic environment where radiated RF disturbances are under con...

Page 13: ...Distributed by CITIZEN is a registered trade mark of Citizen Holdings Co Japan Design and Specification are subject to change without notice ...

Reviews: