AVG
AVG
AVG
AVG
Cuff has been placed
incorrectly.
Worn-out batteries.
SYMBOLS/
SYMPTOMS
CONDITIONS/
CAUSES
INDICATION/
CORRECTION
Note: If "EP" appears on the display, just return the device to your local
distributor or importer.
Shaking of the wrist
with the cuff on.
Air circuit adnormality.
Cuff tube may not be
plugged into monitor
correctly.
Can't determine blood
pressure measurement
data.
Wrap the cuff properly
so that it is positioned
correctly.
Replace them with
2 new AAA (LR03)
alkaline batteries.
Check cuff connection.
Measure again.
Measuring Error
Symbol shown
on display
or the blood
pressure value
is displayed
excessively low
or high.
Inflation pressure
exceeding 300 mmHg.
Did you talk or move
during measurement?
Battery polarities
have been positioned
incorrectly.
Wrap the cuff properly
and keep steady.
Measure again.
Switch the unit off,
then measure again.
Measure again. Keep
wrist steady during
measurement.
Re-insert the batteries
in the correct
positions.
Measuring
Error Symbol
No power-on by
pressing
button
Measuring
Error Symbol
Measuring
Error Symbol