Cirs Dynamic Phantoms 008A User Manual Download Page 7

Removal of Ball Cube Insert (Optional)

1.

CIRS has provided a Ball Cube Removal Tool.  Please note
precautions on Ball Cube insert packaging.

2.

Insert Ball Cube Removal Tool into back of Target Moving Rod
at insertion point as shown above.

3.

Gently push forward until Ball Cube is removed.  Do not use
other tools for this purpose as they may damage the surface of
the Ball Cube which is fragile.

5.

Use a screw driver to tighten the 4 nylon screws on the
cup.

4.

The Model 18023-A and 18043-A rods connect to the motion
actuator with cup holders. Insert the lung rod thru the hole of
the phantom body. If the body is properly aligned the rod will
slip into the cup. Ensure that the pin hole on the rod is aligned
with the pin inside the cup.

Summary of Contents for Dynamic Phantoms 008A

Page 1: ...IMAGE ACQUISITION TREATMENT PLANNING DOSE DELIVERY USER GUIDE Dynamic Phantoms Model 008A 008A 20 18023 A 18043 A 900 Asbury Ave Norfolk Virginia 23513 USA Tel 757 855 2765 WWW CIRSINC COM...

Page 2: ...oval of Ball Cube 7 Model 008A 18023 A 18043 A Cables and Connections 8 4D CT QA Insert Set Up 9 Setup notes for Dynamic Phantoms with controller connecting to a PC using an Ethernet connection 10 CIR...

Page 3: ...rd axis actuator Model 18023 A shares some components with the CIRS Model 008A Dynamic Thorax Phantom As with the Model 008A target and surrogate motion are inde pendently controlled with CIRS Motion...

Page 4: ...en connecting this product to a power source If you do not know whether the outlet is grounded check with a qualified electrician Do not remove ground prong Do not install or use this product near wat...

Page 5: ...he case and set aside Drop N Tell shipping damage indicator shows when a case has been dropped in transit and contains potential dam aged goods The sensor displays a red arrow when applied before ship...

Page 6: ...d is aligned with screw hole on actuator rod 5 Use thumb screw to secure lung rod to actuator rod 6 Push lung rod towards the motor box Visually check gap around the rod It should appear consistent If...

Page 7: ...ll Cube is removed Do not use other tools for this purpose as they may damage the surface of the Ball Cube which is fragile 5 Use a screw driver to tighten the 4 nylon screws on the cup 4 The Model 18...

Page 8: ...ting device to back of actuator Model 008A 18023 A 18043 A Cables and Connections A Adjustable legs kit See B and C for mounting B Prior to attaching phantom body care fully lay base plate with actuat...

Page 9: ...tart angle 0 degree For AP IS 0 LR 0 AP per your choice start angle 90 degree 5 Run phantom and perform 4D CT imag ing Obtain test results from the 4D CT sorted images Report results as the difference...

Page 10: ...d above 1 To install the necessary driver unzip the USB to Network Adapter folder found on the provided USB drive or download the zipped folder from the CIRS website Archive to a known location and se...

Page 11: ...0 249 8 Once the IP address is entered click OK Connect the PC to the Controller using the provided Ethernet cable by inserting one end of the cable in the controller s Ethernet port and the other on...

Page 12: ...optional The software automatically creates a log file where data about wave form parameters are saved The log file is usually located under the current user in the Application Data folder A Windows O...

Page 13: ...g included on provided USB 1 Carry Case OPTIONAL ACCESSORIES INCLUDED WITH MODEL 008A Note Customers must complete their order with the purchase of at least one 1 interchangeable insert option Refer t...

Page 14: ...isfied with your purchase for any reason please contact your local distributor prior to returning the product Visit https www cirsinc com distributors to find your local distributor If you purchased y...

Page 15: ...15 NOTES...

Page 16: ...tice Publication DYNAMIC UG 110320 Computerized Imaging Reference Systems Inc has been certified by UL DQS Inc to ISO 13485 2016 Certificate Registration No 10000905 MP2016 COMPUTERIZED IMAGING REFERE...

Reviews: