Appendix 1
Thermistor Setup:
1.
Position 400 series thermistor center and flush with hose outlet. See diagram below.
Note: Acceptable temperature parameters at hose outlet are as follows:
Low temperature setting: 32.2°C +4.0°C/ -2.0°C (90°F +7.2°F/ -3.6°F)
Medium temperature setting: 37.8°C +4.0°C/ -2.0°C (100°F +7.2°F/ -3.6°F)
High temperature setting: 43.3°C +4.0°C/ -2.0°C (110°F +7.2°F/ -3.6°F)
WarmAir
135 has approval in accordance with ASTM F2196-02 Standards. These standards
are based on contact surface temperature of the blanket being 48°C or less. Temperatures
referenced at the hose outlet do not represent contact surface temperatures of the blanket due
to temperature loss through the hose and dispersion through the blanket.
TYPICAL TEMPERATURE CHARACTERISTICS
TIME
T
EM
PER
A
T
U
R
E
Temperature at Hose Outlet.
Blankets Temperature Range.
Front View
Side View
Extend hose
Thermistor