![Cincinnati Sub-Zero EZT-570S Reference Manual Download Page 28](http://html.mh-extra.com/html/cincinnati-sub-zero/ezt-570s/ezt-570s_reference-manual_2607293028.webp)
EZT-570S User Communication Reference Manual
28
*B12
Parameter
Value
Description
Bit0 Event
1
Bit1 Event
2
Bit2 Event
3
Bit3 Event
4
Bit4 Event
5
Bit5 Event
6
Bit6 Event
7
Bit7 Event
8
Bit8 Event
9
Bit9 Event
10
Bit10 Event
11
Bit11 Event
12
Bit12 Event
13
Bit13 Event
14
Bit14 Event
15
Bit15 Not
Assigned
Example
Turn on the chamber.
According to the event table at the beginning of this section, the chamber event for a standard
temperature/humidity chamber is event 1. The bit number for event 1 is zero, thus the bit at position
(index) zero of the word should be set.
The bit values of the word then become: 00000000 00000001. The decimal equivalent of the binary
array is 1 (0x0001). By setting register 22 to a value of 1, the chamber will turn on.
Example
Turn on customer events 7, 10, 14 and 15.
By comparing the event numbers to their bit positions, set the bits in the word accordingly:
0110001001000000. The decimal equivalent is 25152 (0x6240). Setting register 23 to a value of
25152 will turn on each of the selected customer events.
Setting the individual bits in the word turns on (1) or turns off
(0) the event. If an event is for controlling an option that is not
available on your chamber, the associated bit should be set
off.