High Power DC Electronic Load 63200 Series Operation & Programming Manual
Description Set the load on time of program file.
Syntax
PROGram:ONTime
<NRf>[suffix]
Parameters
30
Sec
Example
PROG:ONT
10
PROG:ONT
100mS
Query
Syntax
PROGram:ONTime?
Return Parameters <NR2> [Unit=Sec]
Query
Example
PROG:ONT?
Return
Example
10
PROGram:OFFTime
Description Set the load off time of program file
Syntax
PROGram:OFFTime
<NRf>[suffix]
Parameters
30
Sec
Example
PROG:OFFT
20
PROG:OFFT
200mS
Query
Syntax
PROGram:OFFTime?
Return Parameters <NR2> [Unit=Sec]
Query
Example
PROG:OFFT?
Return
Example
0.2
PROGram:PF:DELay
Description Set the pass/fail measurement delay time of program file.
Syntax
PROGram:PF:DELay
<NRf>[suffix]
Parameters
60
Sec
Example
PROG:PF:DELay
20
PROG:PF:DELay
200mS
Query
Syntax
PROGram:PF:DELay?
Return Parameters <NR2> [Unit=Sec]
Query
Example
PROG:PF:DEL?
Return
Example
0.2
PROGram:PF:TIME
Description Set the pass/fail measurement time of program file
Syntax
PROGram:PF:TIME
<NRf>[suffix]
Parameters
60
Sec
Example
PROG:PF:TIME
20
PROG:PF:TIME
200mS
Query
Syntax
PROGram:PF:TIME?
Return Parameters <NR2> [Unit=Sec]
Query
Example
PROG:PF:TIME?
Return
Example
0.2
7-22
Summary of Contents for 63200 Series
Page 1: ......
Page 2: ......
Page 10: ......
Page 14: ......
Page 27: ...General Information Model 63205 Model 63206 and 63207 1 13...
Page 30: ......
Page 38: ......
Page 41: ...Operation Overview V MAX 80V Figure 3 2A The Rear Panel of 63201 63202 63203 63204 3 3...
Page 78: ......
Page 122: ......