Rotational Vacuum Concentrator RVC 2-18 CDplus
HCl resistant version
5 Set-up and connection
Version 04/2012, Rev. 2.3 of 03/07/2017
• sb
27 / 84
Translation of the original operating manual
5
Set-up and connection
5.1
Installation site
Operate the RVC only in closed and dry rooms.
• The table must be stable and have a solid, even tabletop.
• Ensure sufficient ventilation. Do not place any paper, cloth, or similar
material behind or under the unit, since otherwise the air circulation will
be impaired.
• Keep a safety distance of at least 30 cm from the wall so that the vents
in the unit remain fully effective.
• The ambient temperature must be in the range of +10°C to +25°C.
• Do not subject the RVC to thermal stress, e.g. by positioning it near
heat generators.
• Avoid direct sunlight (UV radiation).
5.2
Power supply
5.2.1
Connection
DANGER
The operating voltage on the name plate must correspond to the local
supply voltage
CHRIST rotational vacuum concentrators are units of safety class I.Units of
the type
RVC 2-18 CDplus HCl
have a three-wire power cord with an IEC
C13 connector (see chapter 10 - "Technical data").
An equipotential bonding screw is located on the back (see chapter 2.1.1 -
"Functional and operating elements"). This equipotential bonding screw can
be used to perform an earth conductor check.
5.2.2
Customer-provided fuses
Typically, the RVC must be protected with 16 Amp G fuses that are to be
provided by the customer.