Freeze-dryer Alpha 1-4 LDplus
Freeze-dryer Alpha 2-4 LDplus
2 Layout and mode of operation
Version 11/2006, Rev. 2.2 of 17/10/2016
• sb
19 / 102
Translation of the original operating manual
2.2.2.4
Final drying
Final drying is an option whenever one requires a product with minimal
residual moisture. In the physical sense, this process is a desorption
process, i.e. the removal of adsorptively bound water. Final drying is
performed under the lowest possible final pressure that depends on the ice
condenser temperature in accordance with the vapour pressure curve
above ice as well as on the final vacuum of the vacuum pump that is used.
The process is supported by a higher shelf temperature.
Pos : 36 / 010 U ni vers al module/ Leerz eile @ 0\ mod_1202116244500_0. doc x @ 114 @ @ 1
Pos : 37 / 200 C hrist/ 360 GT-BA Labor-Pilot (ST ANDARDM ODU LE)/ 020 Auf bau und Wir kungs weise/020-0020-0020- 0050 Troc knungs ende und Bel üft ung @ 25\ mod_1404983261476_68. doc x @ 183145 @ 4 @ 1
2.2.2.5
End of drying and aeration
The end of the drying process is reached when both the product and shelf
temperature are clearly in the positive range (+15 to +20°C) and if their
difference is not greater than 5 K.
Another indication of the end of the drying process is the behaviour of the
vacuum and of the ice condenser temperature. The ice condenser is no
longer subject to load and reaches the final temperature of approximately -
55°C or -85°C. The pressure in the drying chamber decreases in
accordance with the ice condenser temperature.
The vacuum pump will be switched off and the drying chamber will be
aerated via a rubber valve or via the aeration valve. The aeration valve can
also be used to flood the unit with nitrogen or another inert gas instead of
ambient air.
Then, the product can be removed from the unit.
Pos : 38 / 010 U ni vers al module/ Seitenwec hsel @ 0\ mod_1202116244312_0. doc x @ 105 @ @ 1