ChoiceMMed MD300C228 User Manual Download Page 2

Maintenance and Storage 

1. 

Replace the batteries in a timely manner when low voltage lamp is lighted. 

2. 

Clean surface of the oximeter before it is used in diagnosis for patients. 

3. 

Remove the batteries if the oximeter is not operated for a long time. 

4. 

It is best to store the product in -25

~

+

70

 

and 

≤93% humidity

.

 

5. 

Keep in a dry place. Extreme moisture may affect oximeter lifetime and may cause damage. 

6. 

Dispose of battery properly; follow any applicable local battery disposal laws. 

Cleaning and disinfecting the device 

 

It  is  recommended  to  clean  and  disinfect  the  silicone  touching  the  finger  inside  of  device  with  a  soft  cloth 
dampened with recommended alcohol of 70% isopropyl or 70% ethanol before and after each use. 

 

Excessive disinfection may cause damage to the device and is therefore not recommended for this device 
unless oth

erwise indicated in your hospital’s servicing schedule. 

 

Do not pour or spray liquids onto the device and do not allow any liquid to enter any openings in the device. 
Allow the device to dry thoroughly before reuse. 

Caution: 

Never use EtO (ethylene oxide) or formaldehyde for disinfection. 

The  use  life  of  the  device  is  five  years  when  it  is  used  for  15  measurements  every  day  and  10  minutes  per  one 
measurement. The Oximeter requires no routine calibration or maintenance other than replacement of batteries. 

Specifications 

1. Display Type           

LCD display

 

2. SpO

2

                         

Measurement range: 70%~100% 

 

Accuracy: 70%~100%: ±2%; 

≤69% no definition 

Resolution: 1% 

Note: A functional tester cannot be used to assess the accuracy of a pulse oximeter monitor or sensor. Clinical 
testing is used to establish the SpO

2

 accuracy. The measured arterial hemoglobin saturation value (SpO

2

) of the 

sensors  is  compared  to  arterial  hemoglobin  oxygen  (SaO

2

)  value,  determined  from  blood  samples  with  a 

laboratory CO-oximeter. The accuracy of the sensors in comparison to the CO-oximeter samples measured over 
the  SpO

2

  range  of  70%~100%.  Accuracy  data  is  calculated  using  the  root-mean-squared  (Arms  value)  for  all 

subjects,  per  ISO  80601-2-61,  Medical  Electrical  Equipment

Particular  requirements  for  the  basic  safety  and 

essential performance of pulse oximeter equipment for medical use. 
A  functional  tester  is  used  to  measure  how  accurately  Pulse  Oximeter  is  reproducing  the  specified  calibration 
curve and the PR accuracy. 
The model of functional tester is Index2 FLUKE simulator and the version is 2.1.3.

 

3. Pulse Rate                                       

Measure range: 30bpm~250bpm

 

Accuracy: 30bpm~99bpm, ±2bpm; 100~250bpm, ±2% 
Resolution: 1bpm 

4. Perfusion Index                             

Measure range: 0.3%~20.0% 

5. Probe LED Specifications 

          RED  Wavelength: 660±3nm      Radiant Power: 3.2mw 

 

IR        Wavelength: 905±10nm    Radiant Power: 2.4mw     

 

(NOTE:

 The information about wavelength range can be especially useful to clinicians).

 

6. Power Requirements                   

Two AAA alkaline Batteries

 

Power consumption: Less than 40mA   

7. Environment Requirements     

Operation Temperature: 5

~40

 

Storage Temperature: -25

~+70

 

Ambient Humidity: 15%~93% no condensation in operation;   

≤93% no condensation in storage/transport 

Atmosphere pressure: 70kPa~106kPa 

Notes: When the ambient temperature is 20°C, it is required 6 hours for the equipment to warm from the minimum storage 
temperature or 4 hours to cool from the maximum storage temperature between uses until it is ready for its intended use.

 

8. Equipment Response Time: a

s shown in the following figure, average data update period is 8s.

 

 

9. Classification 

According to the type of protection against electric shock: Internally powered equipment 
According to the degree of protection against electric shock: Type BF applied part (applied part: the rubber hole of 
the device)

 

According to the degree of protection against ingress of water: IP22 
According to the mode of operation: continuous operation 

Clinical Study Summary 

The following details are provided to disclose actual performance observed in the clinical validation study of healthy 
adult volunteers. The ARMS value analysis statement and Bland-Altman plot of data are shown as follows: 

ARMS Value Analysis Statement 

Item 

70--100 

90--100 

80--<90 

70--<80 

70-<90 

#pts 

227 

91 

81 

55 

136.00 

Bias 

0.99 

0.73 

1.44 

0.74 

1.16 

ARMS 

1.80 

1.16 

2.04 

2.21 

2.11 

 

Bland-Altman Plot Graphic 

Possible Problems and Solutions 

 

Problems 

Possible reason 

Solution 

SpO

2

 or PR cannot 

be shown normally. 

1. Finger is not inserted correctly. 
2. 

Patient’s SpO

2

 value is too low to be 

measured. 

1. Retry by inserting the finger. 
2. There is excessive illumination. 
3. Try some more times. If you can make sure 
no problem exist in the product, please go to a 
hospital timely for exact diagnosis. 

SpO

2

 or PR is 

shown unstably. 

1. Finger might not be inserted deep enough. 
2. Excessive patient movement. 

1. Retry by inserting the finger. 
2. Be calm. 

The oximeter 
cannot be powered 
on. 

1. No battery or low power of battery. 
2. Batteries might be installed incorrectly. 
3. The oximeter might be damaged. 

1. Please replace batteries. 
2. Please reinstall the batteries. 
3. Please contact local customer service center. 

Indication lamps 
are suddenly off. 

1. The product is automatically powered off 
when no signal is detected longer than 8s. 
2. The battery power is too low to work. 

1. Normal.   
2. Replace the batteries. 

“Err 7” is displayed 
on screen. 

Err 7 means all the emission LED or 
reception diode is damaged. 

Please contact local customer service center.                                                                     

 

Symbol Definitions 

 

Symbol 

Definition   

Symbol 

Definition   

 

Type BF applied part 

 

Attention 

IP22

 

Protected against dripping water 

%SpO

2

 

Oxygen saturation   

PR bpm 

Pulse rate (BPM) 

 

Low power indication   

 

No SpO

2

 Alarm 

 

Serial No. 

 

Storage temperature and relative 
humidity 

 

Follow instruction for use 

 

Manufacturer’s information 

 

Date of Manufacture 

 

Bluetooth indication   

?

 

Indicate the signal is not stable 

 

Waste electrical and electronic 
equipment 

 

Medical device 

Electromagnetic Compatibility 

The device conforms to IEC60601-1-2:2014 Electromagnetic Compatibility (EMC) standard. 
Essential performance is defined as SpO

2

 accuracy and pulse rate accuracy or an indication of abnormal operation. 

Accuracies may be affected as a result of exposure to electromagnetic disturbances that are outside of the environments 
listed in the intended use. If issues are experienced, move the device away from the source of electromagnetic disturbances. 

Table 1: Electromagnetic Emissions Limits and Compliance 

Emissions Test 

Compliance 

RF Emissions 
CISPR 11 

Group 1, Class B 

Note: Harmonic Emissions (IEC 61000-3-2), Voltage Flicker Emissions (IEC 61000-3-3) are not applicable. 

Table 2: Electromagnetic Immunity 

Immunity Test 

Compliance 

Electrostatic Discharge (ESD) 
IEC 61000-4-2 

±8 kV contact 
±2 kV, ±4 kV, ±8 kV, ±15 kV air 

Rated power Frequency Magnetic Fields 
IEC 61000-4-8 

30 A/m 
50Hz and 60 Hz 

Radiated RF 
IEC 61000-4-3 

80 MHz 

– 2.7 GHz 

10 V/m 80% AM 1kHz 

380 

– 390 MHz 

27 V/m Pulse mod. 18Hz 

430 

– 470 MHz 

28 V/m FM±5Hz deviation 1kHz sine 

704 

– 787 MHz 

9 V/m Pulse mod. 217Hz 

800 

– 960 MHz 

28 V/m Pulse mod. 18Hz 

1.7 

– 1.99 GHz 

28 V/m Pulse mod. 217Hz 

2.4 

– 2.57 GHz 

28 V/m Pulse mod. 217Hz 

5.1 

– 5.8 GHz 

9 V/m Pulse mod. 217Hz 

Note: Electrical Fast Transients (IEC 61000-4-4), Surge (IEC 61000-4-5), Voltage dips (IEC 61000-4-11), Conducted Immunity 
(IEC 61000-4-6) are not applicable. 

 

FCC Declaration 

FCC ID: WWIMD300C228 

Please take attention that changes or modification not expressly approved by the party responsible for compliance could 

void the user’s authority to operate 

the equipment. 

 
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: 
(1) This device may not cause harmful interference, and 
(2) This device must accept any interference received, including interference that may 

cause undesired operation.

 

Box Content 

1. One pulse oximeter        2. One lanyard        3. Two AAA batteries      4. One instruction manual   

Applicable Models 

MD300C208        MD300C228 

Notes:   
1. 

The illustrations used in this manual may differ slightly from the appearance of the actual product. 

2. 

The specifications are subject to change without prior notice. 

Reviews: