Digital Color Doppler Palm Ultrasound System
63
Appendix B: ACOUSTIC OUTPUT REPORT TABLE
Transducer Model: SonoEye P3
Operation Mode: B
Index Label
MI
TIS
TIB
TIC
At
Surface
Below
Surface
At
Surface
Below
Surface
Maximum Index Value
1.10
0.04
0.04
0.04
Index component Value
0.04
0.04
0.04
0.04
Acoustic
Parameters
p
r.α
at
Z
MI
(MPa)
3.09
P
(mW)
0.92
0.92
0.67
P
1*1
(mW)
0.92
0.92
Z
s
(cm)
0.90
Z
b
(cm)
0.90
Z
MI
(cm)
1.15
Z
pii.α
(cm)
1.15
f
awf
(MHz)
7.93
8.20
8.20
10.65
Other
Information
prr
(Hz)
-
srr
(Hz)
9.00
n
pss
NA
1.00
I
pa.α
at
Z
p
ii.α
(W/cm
2
)
437.39
I
spta.α
at
Z
pii.α
or
Zsii.α
(mW/cm
2
)
0.09
I
spta
at
Z
pii
or
Zsii
(mW/cm
2
)
0.09
p
r
at Z
pii
(MPa)
4.15
Operating
control
conditions
Focus
(cm)
1.50
5.50
5.50
1.00
Depth
(cm)
4.90
8.90
8.90
4.90
Freq
MHz
9.00
9.00
9.00
11.00
PRF
HZ
-
-
-
-