15
■
Auto correct
■
Changing alarm temperature [Auto correct]
This is the setting of a reference temperature for judging, Good or Failure, when the body surface
temperature is measured.
[Default: Reference 36.5°C, alarm 38.0°C]
1. Select the [Auto correct] in the [Sensor Setup] menu to display the Auto correct screen.
Press the [Read setup] button at the [Sensor Selection] to read the current settings of the body surface
temperature sensor being connected. These current settings read in are reflected in each set value.
2. Set reference value, alarm value, to set the average frequency
Reference value
:
Replace the value entered by the moving average value. (Set per 0.1 increments)
Alarm value: Sets the value against reference value (moving average value).
(Set per 0.1 increments)
[
Cases]:
If you set Reference values 36.0 , alarm value 38.0 then
Moving average (36.0) for 2.0 (38.0-36.0) deviation alarm is set.
Average frequency: Sets the number of moving average. (3 to 10 times can be set up)
Average high rates : Smaller variation of reference value.
Average low rates : The greater the variability of the reference value.
Please select the optimal value according to operating conditions.
3.
Press
[OK] or [Apply] button to save the settings.
[OK] will be displayed on the below left of the screen and the setting is completed. If the data is not
transmitted, then keep on pressing the [OK] or [Apply] button.
(1)
Auto correct setting changes will clear the
moving average value.
(2)When the first measurement after the moving average value is cleared or factory default setting,
it will always judges failure.
The result is not determined by temperature measurement.
(3) Moving average value not cleared by restarting the power.
(4)Moving average process will not be activated when measuring temperature is less than 25
℃
or more than 45
℃
(Test result will be outputted).