- 114 -
13. Option
13-1. External Operation Settings “Dig Inp”
Using remote contact signal (no-voltage contact: short or open), selection of chart speed or data printing can be executed
without operating keys at the operation/set keys section. To use this function, you need to allocate operation to a terminal
number. Some operations are allocated automatically to specific terminal numbers.
1. Names of the operations executed externally
Operation name
Terminal used
(1) Select chart speed from three speeds
EX1 and EX2 terminals
(2) Message printing (No.01 and 02)
EX1 and EX2 terminals
(3) Message printing (No.01 to 05)
EX1 to EX4 terminals
(4) Data printing
One arbitrary terminal (multiple selection available)
(5) List printing (List No.1, 2 and 3)
One arbitrary terminal (multiple selection available)
(6) Integration value reset
One arbitrary terminal (multiple selection available)
(7) Message printing (No.01 to No.20)
One arbitrary terminal (multiple selection available)
(8) Time correction
One arbitrary terminal (multiple selection available)
2. Operation and terminal contact signal
1) Operations
allocated automatically to specific terminal numbers
ON: short OFF: open
Operation name
Terminal contact signal
(1) Select chart speed
from three speeds
Three chart speeds should be set as well as the setting described in this section.
(See “8-7. Chart Speed Settings”.)
Recording ON/OFF and
chart speed selection
Between COM and EX
terminal
EX1
EX2
Recording ON
CS1 OFF OFF
CS2 ON OFF
CS3 OFF ON
Recording OFF
ON ON
Chart recording should be ON to enable chart speed selection.
(2) Message
printing
(No.01 and 02)
Message should be set as well as the setting described in this section.
(See “8-13. Message Printing 1 Settings”.)
Message No.01
COM and EX1
Message No.02
COM and EX2
A selected message will be printed when a trigger signal (1sec or longer) is given.
Message printing can be executed also by keys.
(3) Message
printing
(No. 01 to 05)
Message should be set as well as the setting described in this section.
(See “8-13. Message Printing 1 Settings”.)
Message
Between COM and EX
terminal
EX1
EX2
EX3
EX4*
No.01
OFF OFF OFF For
trigger
No.02
ON OFF OFF
No.03
OFF ON OFF
No.04
ON ON OFF
No.05
OFF OFF ON
* A selected message will be printed when a trigger signal (1sec or longer) is given after
selecting message No.
Chart recording should be ON to enable message printing.
Message printing can be executed also by keys.
1 sec or more
For trigger
1sec or more