CHINESPORT SPA / BA No.: 02.07.083/01-12 

   

             GB 

- 7 - 

 
 

 
 
 
 
 
 
 
 
 
 
 
 

 
 
 
 
 
 

 
 
 

 

 
 
 
 

      

 

Clean the chair before re-using (see 

care instructions). 

 
 
 
 
 

 

12. Troubleshooting 

 

Fault 

Poss. cause 

Correction 

Armrests do 
not swivel 
properly 

Locking pins are 
caked up with 
deposits 

Clean, check 
periodically 

Locking pins 
cannot be 
pulled out 

Caked up,  
tension spring is 
jammed or 
defective 

Clean, 
replace locking 
pin 

Chair rocks 

Suction caps are 
worn or 
damaged 

Replace suction 
caps 
(always is pairs!) 

Chair rocks 

Legs are not set 
at equal heights 

Check heights 
and adjust as 
necessary 

Adjustable legs 
are jammed. 

Calcium 
deposits, 
gumming up 

Pull out legs and 
clean 

Chair frame 
too unstable 

Screws are 
loose 

Tighten frame 
and castor 
screws 

Chair frame or 
other parts are 
defective 

Overloading, 
wear 

See 
maintenance! 

13. Care  

 

 In general the chair can be cleaned using    

mild 

conventional 

cleaning 

and 

disinfecting  agent.  Next  simply  wipe  the 
chair  with  clean  water  and  cloth,  and  then 
wipe dry with a clean cloth. 

 

Do  not  use  an  abrasive  scouring  agent  or  cleaner 
(read  the  product  instructions!)  that  contain 
aggressive  solvents  (acids,  thinners).  Do  not  use 
stiff  cleaning  brushes,  as  these  will  damage  the 
plastic  surfaces  or  cause  deposits  to  build  up  on 
the plastic parts and other materials. 

 

When  using  the  product  in  a  clinical  area,  the 
shower/commode  chair  must  be  cleaned  on  a 
regular  basis  at  85°C  for  3  minutes  in  a  cleaning 
machine,  in  addition  to  general  cleaning.  First 
remove any loose accessories. 

 
 
 
 

14. Maintenance 
 

The cause of accidents or damage may  be due to 
the late  detection of wear  and/or improper use, as 
well as irregular maintenance. 

 

For  safety  reasons  and  for  preventing 
accidents,  the  chair  must  undergo  a 
complete  maintenance  at  least  once  a 
year.
 (Refer to the Medical Products Act and 
local the MPBetreibV))  

 

For 

documenting 

maintenance, 

REBOTEC 

supplies  a  Maintenance  Book  –  Chairs  (see:  9.). 
Here you will find information on maintenance and 
forms  for  documentation.  You  can  order  these 
User  instructions  or  download  via  our  website  at 
www.rebotec.de.  

 
 

Maintenance extent 
Check at regular time intervals: 
The intervals depend on the frequency the product 
is used and on the condition of the product. 

 

  Check for apparent damage: 

   (cracks, fractures, loose parts). 

  All screwed connections for tightness 

   (especially the frame). 

  Function of the armrests and locking pins. 

  Check accessories for defects. 

 
 

If 

you 

have 

questions 

concerning 

maintenance and repairs to the chair, then 
contact  an  authorized  dealer  or  medical 
equipment supply outlet.  

 
When making repairs, use only original spare parts 
from REBOTEC. Observe all advisories concerning 
guarantee and warranty. 

Adjustment levels 

Summary of Contents for Koln

Page 1: ...mber Shower Chair Würzburg Art No 01816 130 kg height adjustable Serial Number Shower Commode Chair Ulm Art No 01820 130 kg height adjustable Serial Number Important advisories Read this User Instructions before first time use of the chair and keep it at hand for the user Adhere to all advisories especially the safety advisories Keep this User Instructions booklet at hand for future use and subseq...

Page 2: ... 3 1 2 Intended proper use advisories 3 1 3 Shower chair 3 1 4 Commode chair 3 1 5 Shower Commode chair 3 2 Warranty 3 3 Spare parts orders complaints 3 4 Re use lifetime 3 5 Disposal 3 6 Symbol advisory 3 7 Safety advisories use 4 8 Product specifications 4 8 1 Scope of delivery 4 8 2 Technical specifications 4 9 Spare parts accessories 5 10 Setting up the chair for use 5 11 Using the chair 5 11 ...

Page 3: ...e is a 2 year warranty on our shower and commode chairs A prerequisite for the warranty is use in accordance to the requirements outlined herein as well as conducting regular maintenance Excluded are all wear parts Not observing the User Instructions improper maintenance unauthorised modifications done to the product without the manufacturer s permission automatically lead to the immediate nullifi...

Page 4: ...ach between the seat and the frame when pressing the seat into the seat frame Danger of falling Sitting down standing up and moving the patient should be supervised by a qualified care worker as to the health condition of the patient Use of the chair by patients who do not have complete physical control and normal movement must be done with supervision Warning Danger of Injury At surface temperatu...

Page 5: ...ker of the lower frame 2 Swivel the armrests forwards until they click into the slots of the backrest and the locking pins click into the lower frame Be sure they are securely in place 3 For chairs with height adjustment the seat height of the chair is set at its lowest position for delivery Base setting Adjust the height as necessary see 11 7 11 Using the chair When using the chair the following ...

Page 6: ...can be removed from its receptacles on the chair frame for the purpose of cleaning or exchanging simply by pulling it up 11 5 Removing the seat back cushion The back cushion is fastened to the cross brace or backrest with studs Press the buttons to remove the back cushion 11 6 Handling the commode pan Push the commode pan under the seat from behind the chair To remove the pan lift it up slightly a...

Page 7: ...nd other materials When using the product in a clinical area the shower commode chair must be cleaned on a regular basis at 85 C for 3 minutes in a cleaning machine in addition to general cleaning First remove any loose accessories 14 Maintenance The cause of accidents or damage may be due to the late detection of wear and or improper use as well as irregular maintenance For safety reasons and for...

Page 8: ... maintenance or repair of the chair is done we recommend a complete cleaning for assuring hygiene Follow the cleaning instructions Accessories Armrest cushion Armrest cushions can be clamped onto the armrests retroactively on both the right and left sides The armrest cushion allows better handling when administering infusions or taking blood samples Weight 0 250 kg Manufacturer REBOTEC Rehabilitat...