Clean the chair before re-using (see
care instructions).
12. Troubleshooting
Fault
Poss. cause
Correction
Armrests do
not swivel
properly
Locking pins are
caked up with
deposits
Clean, check
periodically
Locking pins
cannot be
pulled out
Caked up,
tension spring is
jammed or
defective
Clean,
replace locking
pin
Chair rocks
Suction caps are
worn or
damaged
Replace suction
caps
(always is pairs!)
Chair rocks
Legs are not set
at equal heights
Check heights
and adjust as
necessary
Adjustable legs
are jammed.
Calcium
deposits,
gumming up
Pull out legs and
clean
Chair frame
too unstable
Screws are
loose
Tighten frame
and castor
screws
Chair frame or
other parts are
defective
Overloading,
wear
See
maintenance!
13. Care
In general the chair can be cleaned using
a
mild
conventional
cleaning
and
disinfecting agent. Next simply wipe the
chair with clean water and cloth, and then
wipe dry with a clean cloth.
Do not use an abrasive scouring agent or cleaner
(read the product instructions!) that contain
aggressive solvents (acids, thinners). Do not use
stiff cleaning brushes, as these will damage the
plastic surfaces or cause deposits to build up on
the plastic parts and other materials.
When using the product in a clinical area, the
shower/commode chair must be cleaned on a
regular basis at 85°C for 3 minutes in a cleaning
machine, in addition to general cleaning. First
remove any loose accessories.
14. Maintenance
The cause of accidents or damage may be due to
the late detection of wear and/or improper use, as
well as irregular maintenance.
For safety reasons and for preventing
accidents, the chair must undergo a
complete maintenance at least once a
year. (Refer to the Medical Products Act and
local the MPBetreibV))
For
documenting
maintenance,
REBOTEC
supplies a Maintenance Book – Chairs (see: 9.).
Here you will find information on maintenance and
forms for documentation. You can order these
User instructions or download via our website at
www.rebotec.de.
Maintenance extent
Check at regular time intervals:
The intervals depend on the frequency the product
is used and on the condition of the product.
•
Check for apparent damage:
(cracks, fractures, loose parts).
•
All screwed connections for tightness
(especially the frame).
•
Function of the armrests and locking pins.
•
Check accessories for defects.
If
you
have
questions
concerning
maintenance and repairs to the chair, then
contact an authorized dealer or medical
equipment supply outlet.
When making repairs, use only original spare parts
from REBOTEC. Observe all advisories concerning
guarantee and warranty.