6 Measurement Optimization
6.11 Rapid Data Transmission
205
Click the
[Channel]
box in the
trigger setting
zone and select
Channel 1
.
Click the
[Continuous]
radio box in the
trigger setting
zone.
3) Set the trigger mode of Channel 2: update measurement data when required.
Click the
[Channel]
box in the
trigger setting
zone and select
Channel 2
.
Click the
[Single]
or
[Groups]
radio box in the
trigger setting
zone. If the
group triggering
mode is selected,
enter the group number in the
[Groups]
input box.
4) Click
[Soft keyboard]
in the
[System]
menu. The [Soft keyboard] dialog box will appear. Click
[Stimulus] >
[Trigger]
in the dialog box and the toolbar to trigger the corresponding soft key will appear.
5) Click the window below to set Channel 2 as the current active channel.
6) Update the measurement data of Channel 2.
7) Click the
[Single]
,
[Groups]
or
[Re-sweep]
on the
trigger
soft key toolbar. Measurement data of Channel 2 will
be updated.
6.10.2 Automatic Change of Measurement Setting
Automatically change measurement settings by means of programming to effectively improve the efficiency of
multi-parameter measurements.
6.10.3 Rapid recalling of measurement
The most effective method to recall the measurement is to save a group of measurements as one instrument state.
➢
The time difference of recalling the instrument with one or more measurements is little.
➢
The corresponding time is required in each recalling. The recalling time can be saved by setting a group of
measurements.
6.11 Rapid Data Transmission
Rapid data transmission can help to improve the measurement efficiency. You can apply the following methods to
increase the data transmission velocity.
●
Rapid recalling of measurement
··················································································
●
Minimization of transmission data
················································································
●
Use of real number format
························································································
●
Use of LAN
●
Use of COM program
······························································································
6.11.1 Use of “Single” Trigger Mode
Use the “single” trigger mode to complete measurement before data transmission.
Menu path:
[Stimulus] > [Trigger] > [Trigger...]
. The
[Trigger
] dialog box will appear.
Select the
[Internal]
radio box in the
trigger source
zone.
Select the measurement channel for data transmission by clicking the
[Channel]
box in the
trigger setting
zone.
Click the
[Continuous]
radio box in the
trigger setting
zone and then click
[OK]
to close the dialog box.
Update the measurement results:
1) Click the corresponding trace state button to set the trace and channel to be updated as the active trace and
channel.
2) Click
[Soft keyboard]
in the [System] sub-menu. Then click
[Stimulus] > [Trigger]
in the
[Soft keyboard]
dialog box.
Summary of Contents for AV3672 Series
Page 3: ......
Page 4: ...AV3672 Series Vector Network Analyzer Contents...
Page 5: ......
Page 124: ...5 Menu 5 1 Menu structure 120 5 1 2 Track Fig 5 2 Track Menu...
Page 125: ...5 Menu 5 1 Menu structure 121 5 1 3 Channel Fig 5 3 Channel Menu...
Page 126: ...5 Menu 5 1 Menu structure 122 5 1 4 Excitation Fig 5 4 Excitation Menu I...
Page 127: ...5 Menu 5 1 Menu structure 123 Fig 5 5 Excitation Menu II...
Page 128: ...5 Menu 5 1 Menu structure 124 Fig 5 6 Excitation Menu III...
Page 129: ...5 Menu 5 1 Menu structure 125 5 1 5 Response Fig 5 7 Response Menu I...
Page 130: ...5 Menu 5 1 Menu structure 126 Fig 5 8 Repsonse Menu II...
Page 131: ...5 Menu 5 1 Menu structure 127 Fig 5 9 Response Menu III...
Page 132: ...5 Menu 5 1 Menu structure 128 Fig 5 10 Response Menu V Fig 5 11 Response IV...
Page 133: ...5 Menu 5 1 Menu structure 129 5 1 6 Calibration Fig 5 12 Calibration Menu...
Page 134: ...5 Menu 5 1 Menu structure 130 5 1 7 Marker Fig 5 13 Cursor Menu I...
Page 135: ...5 Menu 5 1 Menu structure 131 Fig 5 13 Cursor Menu II...
Page 136: ...5 Menu 5 1 Menu structure 132 Fig 5 15Marker Menu III...
Page 137: ...5 Menu 5 1 Menu structure 133 5 1 8 Analysis Fig 5 16 Analysis Menu I...
Page 138: ...5 Menu 5 1 Menu structure 134 Fig 5 17 Analysis Menu II...
Page 139: ...5 Menu 5 1 Menu structure 135 Fig 5 18 Analysis Menu III...
Page 140: ...5 Menu 5 1 Menu structure 136 5 1 9 System Fig 5 19 System Menu I...
Page 141: ...5 Menu 5 1 Menu structure 137 Fig 5 20 System Menu I...
Page 254: ...8 Basis of Network Measurement 8 3 Amplifier Parameter Specifications 250...
Page 257: ...8 Basis of Network Measurement 8 4 Complex Impedance 253...
Page 373: ...Appendix Appendix 4 Pulse Measurement 369 Fig 4 9 Receiver gain configuration Dialog Box...