HOIST 

100 Series

Page 8

Issue 

2.0

June 2018

Troubleshooting

Further assistance

Please telephone Chiltern Invadex on 01869 365500 further assistance.  Remember that service calls made 

after the warranty period will be chargeable.  If the maintenance of the equipment is the responsibility of a 

third party, for example the Social Services, please contact them in the first instance.

Battery models only

Hoist lacking power

Battery is low

Charge batteries

Hoist make beeping sound

Battery is low

Hoist is left off charge

Charge batteries

Return to charge point

Red light is on

Battery is low

Charge batteries

Red light is flashing

Batteries are critically low

Charge immediately

Green light is on

Batteries charging

Will turn off when charged

Red & Green lights are 

flashing

Hoist idle

Return to charge

Hoist will not function

Hoist is charging

Remove hoist from charging dock and 

wait for green LED to turn off

Powered models only

No hoist function

No power

Ensure mains power is connected,  

check fuse box

All models

Hoist will not function

Hoist has been overloaded

Emergency stop has been 

activated

Thermal fuse requires reset see page 

3

Reset hoist, see page 5

Hoist stops when lifting

Tape is twisted

Ensure tape is vertical

Orange light is on

Service is due

Arrange service call

Orange light is flashing

Service is overdue

Arrange service call

Hoist lifts when down button 

is pressed

Airline incorrectly installed

Lifting tape may be reversed 

on the spool

Black line on handset must align with 

the raised edges on the hoist connec-

tion when inserted into the grommet on 

the hoist box

To correct press up button to lower 

carry bar whilst pulling lift tape down 

gently to the end of its travel. Tape will 

then rewind itself correctly, or the lower 

cutout will operate.  If lower cutout 

operates the tape is wound correctly.

Symptom                                    Suggested cause                      Solution

Summary of Contents for WISPA 100 Plus

Page 1: ...100 Plus models WISPA 100 Plus This user guide contains essential information in relation to safe use and troubleshooting and should be retained with the hoist for reference Unit 6C Thorpe Drive Banbu...

Page 2: ...Chiltern Invadex hoist has been designed and tested to meet the highest standards of safety Read the safety section of this guide prior to use this is contained on pages 1 2 Intended use 1 Safety cons...

Page 3: ...oad of the sling or any other lifting accessory where this differs from the hoist the lowest weight must not be exceeded When using the Hoist Ensure when the hoist is in use that the surrounding area...

Page 4: ...lly or where there is a change in user circumstance property or equipment ISO 10535 recommends that a conveniently placed telephone or alarm system should be provided as a safety control measure As it...

Page 5: ...Button In event of overload press button to re set Product Identification Label serial number model SWL Carry Spreader Bar Lifting Tape Handset Cable Retaining Socket Emergency Stop Emergency Lower So...

Page 6: ...ction is controlled by the bottom buttons on the handset The buttons are colour coded with red and green These colours correspond to the red and green direction travel arrows on the underside of the t...

Page 7: ...gency lowering pull cord Emergency Stop To operate in case of an emergency pull the red cord to switch off the hoist completely E M E R G E N C Y S T O P L O W E R Emergency stop pull cord RESET Reset...

Page 8: ...en the hoists batteries are in need of charging This is unlikely to happen unless the hoist is used for long periods of time approximately 60 lifts without charging or the hoist has not been properly...

Page 9: ...at point re installation Information regarding LOLER requirements can be seen at http www hse gov uk pubns hsis4 pdf Maintenance of this equipment should only be conducted by engineers trained and ce...

Page 10: ...is charging Remove hoist from charging dock and wait for green LED to turn off Powered models only No hoist function No power Ensure mains power is connected check fuse box All models Hoist will not...

Page 11: ...rack to bottom of carry bar hook at max height is 442mm This is adjustable and should be set by the installer to meet the requirements of the user Typically the carry bar would stop about 300mm from t...

Page 12: ...ation Notes Important Information Battery Powered Models Where this hoist is to be a replacement unit on an existing track installation the charging unit included in the box MUST be used This hoist is...

Page 13: ...arry bar can swivel smoothly care should be taken to ensure that the PTFE bearings are flat against the top of the carry bar Place the retainer housing over the lift tape 5 and insert the dowel throug...

Page 14: ...Invadex UK Limited shall be used to repair this product Use of unauthorised parts will invalidate all warranties and remove all liability from Chiltern Invadex UK Limited for the safety of this produ...

Page 15: ...s Engineer name Comments Year 2 Date of Service Load tested to kg Serviced by Address Engineer name Comments Year 3 Date of Service Load tested to kg Serviced by Address Engineer name Comments Wispa H...

Page 16: ...iced by Address Engineer name Comments Year 6 Date of Service Load tested to kg Serviced by Address Engineer name Comments Year 7 Date of Service Load tested to kg Serviced by Address Engineer name Co...

Page 17: ...ad tested to kg Serviced by Address Engineer name Comments Year 10 Date of Service Load tested to kg Serviced by Address Engineer name Comments In addition to regular service and maintenance a require...

Page 18: ...The below space can be used to store details of your service contractor if the equipment has been provided by a third party for example the Local Authority this information will be available from the...

Page 19: ...odels WB WTB WLB WLTB 200L 200LT 200LTR 200LTX 200LTXR 300L 300PT 300PTR 300P 300PR MHWMEHUKEL We hereby certify that the products listed have been designed assessed manufactured labelled and packaged...

Page 20: ...y disposed of please contact your Local Authority the dealer from where the product was purchased or Chiltern Invadex UK Limited WEEE producer number WEE JD1274RU To request a copy of this guide in la...