9 Technical specifications
9.1
The NucleoCounter
Specificity
The NucleoCounter counts mammalian cell
nuclei stained with the DNA specific
fluorescent dye, propidium iodide.
Sample consumption
100 µl is recommended for NucleoCassette
loading. Approximately 60 µl is loaded into
the cassette.
Analysis volume
Approximately 2 µl of the sample loaded
into the NucleoCassette is transported to
the chamber and is analyzed.
Measurement range
The measurement range is 5x10
3
– 2x10
6
cells/ml.
The optimal measurement range is 1x10
5
–
2x10
6
cells/ml.
Analysis time
When pressing “Run” on the NucleoCounter
the result will be displayed within 30
seconds.
Operation
Menu-controlled by means of keyboard and
LCD display.
Physical data
Weight
Height
Width
Depth
3 kg
26 cm
38 cm
22 cm
Power
See Section 10.2 Power Supply
Input
Fuse
12VDC (11-13VDC)
UL Listed fuse:
600mA/250Vac (UL 248)
Power consumption
NucleoCounter
Peak
Ready mode
Standby
25 W
2.5 W
2 mW
Operation conditions
Maximum relative humidity 80 percent for
temperatures up to 31°C decreasing linearly
to 65 percent relative humidity at maximum
35°C; minimum temperature 15°C.
USB
USB, version 1.1.
Note: Does not support USB Hubs.
Printer
Thermal or Impact printer with RS-232
interface and cable supplied by ChemoMetec
A/S.
Other environmental operating conditions for the NucleoCounter
Indoor use only.
Summary of Contents for NucleoCounter
Page 1: ...NucleoCounter User s Guide P N 991 0009 Revision 1 4 Technology that counts...
Page 2: ...i This page was intentionally left blank...
Page 4: ...iii This page was intentionally left blank...
Page 6: ...v This page was intentionally left blank...
Page 8: ...vii This page was intentionally left blank...
Page 10: ...ix This page was intentionally left blank...
Page 13: ...Introduction and intended use xii This page was intentionally left blank...
Page 21: ...1 Inspection and Unpacking of Equipment 4 This page was intentionally left blank...
Page 28: ...Fejl Henvisningskilde ikke fundet 11 This page was intentionally left blank...
Page 42: ...5 Operation of the NucleoCounter 25 This page was intentionally left blank...
Page 49: ...6 Total and viability count 32 This page was intentionally left blank...
Page 53: ...7 Maintenance of NucleoCounter 36 This page was intentionally left blank...
Page 58: ...8 Troubleshooting Error messages 41 This page was intentionally left blank...