9 Technical specifications
37
9 Technical specifications
9.1
Technical specifications for NucleoCounter YC -100
Specificity
Cells stained with the DNA specific fluorescent dye, PI.
Sample types
Suspension of dispersed yeast cells.
Sample consumption
A minimum of 100µl is recommended for the NucleoCassette
loading. Approximately 50 µl of sample is loaded into the
cassette.
Analysis volume
Approximately 1 µl of the sample is being analyzed in the
NucleoCounter YC-100.
Measurement range
Measurement range:
Optimal Measurement range:
5x10
3
to 6x10
6
cells/ml
5x10
4
to 6x10
6
cells/ml
Presentation of Result
The NucleoCounter YC-100 presents the estimated cell
density in the sample in the cassette. In order to obtain the
cell concentration in the original ce ll suspension the result
must be multiplied with the appropriate multiplication factor
to compensate for any dilution.
Operation
Menu-controlled by means of keyboard and LCD display.
Analysis time
When pressing “
Run
” on the NucleoCounter YC-100 the
result will be displayed within 30 seconds.
Physical data
Weight
Height
Width
Depth
3 kg
26 cm
38 cm
22 cm
Power
See Section 10.2 Power Supply
Input
Fuse
12VDC (11-13VDC)
UL Listed fuse:
600mA/250Vac (UL 248)
Power consumption
NucleoCounter
Peak
Ready mode
Standby
25 W
2.5 W
2 mW
Operation conditions
Maximum relative humidity 80 percent for temperatures up
to 31°C decreasing linearly to 65 percent relative humidity at
maximum 35°C; minimum temperature 15°C.
USB
USB, version 1.1.
Note: Does not support USB Hubs.
Printer
Thermal or Impact printer with RS-232 interface and cable
supplied by ChemoMetec A/S.
Other environmental operating conditions for the NucleoCounter
Summary of Contents for 900-0300
Page 1: ...NucleoCounter YC 100 User s Guide Revision 1 3 Technology that counts...
Page 2: ...i This page was intentionally left blank...
Page 4: ...iii This page was intentionally left blank...
Page 6: ...v This page was intentionally left blank...
Page 8: ...vii This page was intentionally left blank...
Page 10: ...ix This page was intentionally left blank...
Page 12: ...Introduction and intended use xi This page was intentionally left blank...
Page 20: ...1 Inspection and Unpacking of Equipment 8 This page was intentionally left blank...
Page 26: ...Fejl Henvisningskilde ikke fundet 14 This page was intentionally left blank...
Page 44: ...7 Maintenance of NucleoCounter 32 This page was intentionally left blank...
Page 48: ...8 Troubleshooting Error messages 36 This page was intentionally left blank...
Page 55: ...43 2 1 3 Metal E nclosure Figure 36 Printer Output connector seen from the cable entry side...