www.cglifesciences.com
Step 3: While holding the stirrer shaft, slide the mechanical seal stir
bearing over the end of the shaft (a small amount of mineral oil can
be applied to the stirrer shaft to help), through the bottom of the
bearing. Once the bearing is installed in the center neck, continue to
pull the shaft through the bearing and adjust to proper height.
Please note: The stirrer shaft should be passed thru the bearing
spindle approximately 2” past the final installation position (towards
the motor) and then pushed downward to ensure the face seals
remain in contact with each other.
SHAFT NOTES: BEARING IS DESIGNED FOR USE WITH ROUNDED OR
BEVELED SHAFT ENDS. DO NOT USE WITH FLAT ENDED SHAFTS.