7- Click on the
Settings
tab.
8- Select
Auto Detect
from the
Emulation
pull down menu.
9- Click on the
ASCII Setup
button.
10- Check the
Echo typed characters locally
check box.
11- Click
OK.
12- Click
OK.
Now if the communication port is configured correctly the displayed value on the controller should appear
in the HyperTerminal window. The controller is set to output a value every 10 seconds; this value can be
adjusted if needed.
Adjusting the output time duration:
1- Depress the Program and Down button on the display window.
2- Press the Program Button repeatedly until the display toggles between [SPC_5] and [000].
3- Using the Up or Down buttons adjust the display to read [007].
4- Press the Program button until [on_t] and [00010] is displayed.
5- Using the Up or Down buttons adjust the display to read the desired output time in seconds.
6- Press the Program button.
7- Depress the Program and Down button.
Analog Output:
(A 3-pin Nano connector is required. P/N: CG-2033-50)
The analog outputs the Forward (CW) RPM value only
. The analog is set to output 10Vdc when the
contr
oller’s speed is at 100% and 0Vdc when set 0%. To approximate the analog output value per volt dc,
divide the max rpm display by 10.
This will give you rpm’s per 1-volt dc.
Example:
Max RPM: 500
500/10= 50
Therefore every 50
rpm’s = 1Vdc
In this example: If 90
rpm’s are displayed, then the analog output is about 1.8 Vdc
(90/50 = 1.8 Vdc)
Note: Analog output will not work properly when the motor is reversed or when the motor controller is
switched to Torque. To display torque or the motor’s rpm’s in reverse use the Serial output.
1- Attach the yellow analog cable to the Analog port located in the rear of the motor controller.
2- Attach the brown wire to the (+ Vdc) terminal on your analog input device.
3- Attach the blue wire to the (- Vdc) terminal on your analog input device.
Warranty
Chemglass, Inc. guarantees this unit against defects in material and workmanship for a period of two
years from the date of sale. If the unit should malfunction it must be returned for evaluation. If the unit is
determined to have a defect in materials or workmanship then it will be repaired or replaced at no charge.
Tampering or damage resulting from excessive current, heat, moisture, vibration, corrosive materials, or
misuse will void this warranty. Programming changes or reconfiguration are not covered under warranty.
To return this item under warranty, advise Chemglass technical service department of the situation and
they will provide you with a return authorization number. The unit will be evaluated and repaired as
required in accordance to the warranty.
Please contact the Chemglass technical service department with warranty repair
information.