The display will continue to show the real time gas reading whether the horn is acknowledged or not. 

 
Latch

  

Enables configuration of the low level alarm and the High Level Alarm to latching or non-latching 
status. 

Latching

 means that once the alarm level is reached, the system will remain in alarm status even 

if the gas concentration subsequently drops below the alarm level. In 

non-latching

 mode, the alarm 

status goes away if the gas concentration drops below the alarm level. See Acknowledge section above 
for further explanation. 
 

Altitude 

The response of most electrochemical sensors is proportional to absolute concentration (mg/l) and not 
the relative concentration (% vol., ppm or ppb). The sensitivity, of the sensor, therefore depends on the 
altitude of the monitor since the sensitivity is a function of atmospheric pressure. The altitude is usually 
a one-time configuration setting that is made at the time of installation of the monitor. If the monitor is 
not moved to a vastly different location, then the altitude need never be changed again. If the altitude is 
not set, then the monitor will use the default value corresponding to sea level. The altitude can be 
entered in feet.  

 
CHANGING CONFIGURATION SETTINGS USING THE MAGIC BUTTON™ 

You can use the magic button to access the area monitor’s configuration menu. The following settings 
can be changed using the magic button: Low alarm setting, high alarm setting, horn volume, and zero 
adjust. 
 

Adjusting the Low Alarm Setting Using the Magic Button: 

Use this adjustment to choose the gas level in parts per million (ppm) when the low alarm will sound.  
 
Note: Once you enter the configuration menu, if the monitor detects no button presses for five seconds, 
it will exit the menu and return to the main display.  
 

1.

 

To enter the configuration menu, 

press and hold

 the Magic Button™ until you see MENU on the 

main display. Then 

release

 

the button. 

2.

 

Press and hold

 the button to cycle through the sub menus. The display will repeat cycling 

through the sub menus as long as you hold down the button. When you see LALM displayed, 

release

 the button.  

3.

 

Press and hold

 the button. The alarm setting will increase in value. 

Release

 the button when you 

see the value you want. (To alternate between increasing or decreasing the setting, 

release the 

button and then press and hold again

.) 

4.

 

If the monitor detects no button presses for five seconds, the monitor will exit the configuration 
menu, and then it will display in sequence the low alarm setting, high alarm setting, and the horn 
volume. The monitor will then return to the main display. Your alarm level adjustment has been 
saved. 

 

Adjusting the High Alarm Setting Using the Magic Button: 

Use this adjustment to choose the gas level in parts per million (ppm) when the high alarm will sound.  

Summary of Contents for Steri-Trac

Page 1: ...Document Number CM 112 A 0 0 5 Copyright ChemDAQ Inc 2019 All Rights Reserved ChemDAQ Steri Trac Gas Monitor Operation Manual...

Page 2: ...he purchaser all warranties of manufacturers of such components THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES EXPRESSED IMPLIED OR STATUTORY AND IS STRICTLY LIMITED TO THE TERMS HEREOF SELLER SPECI...

Page 3: ...3...

Page 4: ...Trac monitor can be used as a standalone instrument the full benefit of the design is only achieved when the monitor is integrated into a gas monitoring system A full gas detection system consists of...

Page 5: ...hemDAQ s Installation Responsibilities 15 Installation of Steri Trac and PointGard II Monitors in Same SystemError Bookmark not defined CONFIGURATION OF MONITOR 15 Alarm High Level Adjust 15 Alarm Low...

Page 6: ...displays The system has been designed to be fail safe so that in the unlikely event that one component were to fail the rest of the system will continue to operate normally For example the DAQ and all...

Page 7: ...ss the danger from outside a room before entering Use of the remote display is very simple The remote display is wall mounted connected by cable to the monitor and in some cases connected to a local p...

Page 8: ...ted to ChemDAQ s proprietary data acquisition system DAQ The monitors can also be configured through the DAQ computer See discussion below and DAQ manual for further information Remote Video Display T...

Page 9: ...ertain gases ChemDAQ s Spot On Filter that makes the sensor much more specific to the target gas which results fewer false alarms This proprietary filter removes common interferents which for example...

Page 10: ...C supply at 50 60 Hz and provides power to the instrument together with up to three remote displays Major Components of the Steri Trac Monitor Display and Horn Display Four digit ultra bright 5x8 Tri...

Page 11: ...all box is to be mounted The wall box is designed to be hard wired to the electrical supply but it can be plugged into a regular outlet by means of an optional cable It is recommended that the power s...

Page 12: ...tor with a customer s instrument panel or for activation of external ventilation Connection to these relays is made via the terminal block in the Monitor Wall Box as shown in Figure 1 above Remote Dis...

Page 13: ...minal block connections in the Remote Display J2 1 3 4 5 6 7 8 2 Gnd V Tx Tx Rx Rx Blue Indicates Signal Wiring RS 485 Red wires power 15 VDC Connect Rx To Area Montor Rx to Disp Rx to Disp Tx Connect...

Page 14: ...surfaces are unsuitable for screw attachment or a vertical mounting surface is not available For these applications the Remote Sensor can be attached to flat horizontal surfaces with Velcro 2 or simil...

Page 15: ...el alarm set point The high level alarm must be higher or equal to the low level alarm Alarm Low Level Allows you to adjust the low level alarm set point Horn Volume The horn volume is adjustable to a...

Page 16: ...area monitor s configuration menu The following settings can be changed using the magic button Low alarm setting high alarm setting horn volume and zero adjust Adjusting the Low Alarm Setting Using th...

Page 17: ...horn volume on a scale from 1 lowest volume to 8 highest volume The horn will sound as the value changes Note Once you enter the configuration menu if the monitor detects no button presses for five s...

Page 18: ...monitor will go through its zero adjust function and will flash in green Z0 0 4 After a few seconds the monitor will display in sequence the low alarm setting high alarm setting the horn volume then i...

Page 19: ...y high concentrations of the target gas or high concentrations of other gases or vapors then the sensor should be recalibrated Non routine calibration can be arranged through ChemDAQ s customer servic...

Page 20: ...Service Repair 300 Business Center Drive Pittsburgh PA 15205 Emergency Response ChemDAQ is not able to recommend specific actions to be taken in the event that the ChemDAQ Gas Monitoring System initia...

Page 21: ...e remedial action There is an interferent gas in the air Filter needs to be replaced contact ChemDAQ customer service Small signals absent target or interferent gas can be due to sensor drift Adjust z...

Page 22: ...22 SPECIFICATIONS...

Page 23: ...F 20 to 50 o C 4 F to 122 F 20 to 50 o C 4 F to 122 F Pressure Range steady state Atmosphe ric 10 Atmospheric 10 Atmospheric 10 Atmospheric 10 Atmospheric 10 Atmosphe ric 10 Response Time T90 s 140 1...

Page 24: ...4 Remote Display Output RS485 Supply Voltage 100 240 VAC 75A 50 60 Hz Hot Swap Sensor Module Yes Remote Mount Sensor Module Capability Yes Local PC Connection serial port on area monitor RS 232 RS 485...

Page 25: ...ally Safe Area Monitor for Hydrogen Peroxide ST2 ISHMON Steri Trac Intrinsically Safe Area Monitor for Ozone ST2 ISOMON Steri Trac Intrinsically Safe Area Monitor for LEL ST2 LELMON Steri Trac Remote...

Page 26: ...Remote Sensor Wiring Detail V Sens Sens Gnd 1 3 4 2 Blue Indicates Signal Wiring RS 485 Thick Red Indicates power wires 15VDC Remote Sensor Wiring Diagram Power Signal cable available from ChemDAQ To...

Page 27: ...Connections are for Daisy Chain Option of Remote Displays Tx from First Display to Rx of second in chain Tx from First Display to Rx of second in chain Remote Display Wiring Area Monitor Disp V Disp...

Page 28: ...Business Center Drive Pittsburgh PA 15205 Phone 412 787 0202 Toll Free 800 245 3310 Fax 412 788 2526 For a General Informational Question Please e mail info ChemDAQ com For a Technical Support Request...

Reviews: