Page 4
Section 3 -
Installation and Interconnections
The instrument is designed to be panel mounted, and is a standard 96x48mm panel meter
enclosure. Note that the instrument must always be situated in such a way as to enable
adequate air circulation about the unit.
If a transducer was supplied with the CCD100 a suitable cable may have been included.
3.1
Transducer socket Pinout - 15 Way 'D' Type
Pin Number
Designation
2
Signal
5
Power Gnd
6
-15V Supply
7
+15V Supply
8
Command Output (V)
or Excitation
(mV option)
11
Signal Return
12
Signal Return
13
+24V Supply
15
Earth/Chassis
All other pins
Not Connected
If the transducer is being powered from another source (e.g. a mains powered supply) it
will only be necessary to connect to the two signal pins.
Note: The CCD100 has a fully floating differential input. For single ended use (i.e.
tranducers with differential output) join 0V (pin5) & Signal Return (pin12) at the transducer.
Ensure the common mode input range is not exceeded.
3.2
Interface Socket Pinout – Miniature 15 Way 'D' Type
Pin Number
Designation
1
Analogue Output
2
Relay 2 Com
3
Relay 2 NO
4
Relay 2 NC
5
0V (Signal Ground)
6
External Setpoint Return
7
N/C
8
External Setpoint
9
N/C
10
0V (Signal Ground)
11
TTL Input – Disable Menu
12
TTL Input – Disable Zero
13
Relay 1 Com
14
Relay 1 NO
15
Relay 1 NC