![Chauvin Arnoux C.A 757 User Manual Download Page 14](http://html1.mh-extra.com/html/chauvin-arnoux/c-a-757/c-a-757_user-manual_2594600014.webp)
25
3.3. ENVIRONMENTAL CONDITIONS
Operating range:
-10°C at 55°C and
≤
80%RH without condensation up to
40°C.
Storage range (without battery):
-20°C at +55°C and
≤
90%RH without condensation up to
45°C.
If an extended period of non-use is anticipated, or for storage,
withdraw the batteries from the housing.
For use indoors and outdoors without rain.
Pollution degree: 2.
Altitude: <2000m.
3.4. POWER SUPPLY
The instrument is powered by two 1.5V alkaline batteries
(type AAA or LR3).
Battery life is 100 h.
3.5. CHARACTERISTICS OF CONSTRUCTION
C.A 757
Dimensions (L x l x P) 180 x 52 x 45 mm
Mass
200 g approx.
Cable
length 142 cm
Protection rating
IP 54 according to IEC 60529
IK 04 according to IEC 50102
Drop test
2 m.
MiniFlex sensor
Clamping diameter
70 mm
Length of the sensor
250 mm
Length of the connecting cable 1m, terminated by a specific
three-point plug
Mass
6,0 g approx.
Protection rating
IP 50 according to IEC 60529
IK 04 according to IEC 50102
3.6. ELECTRICAL SAFETY
Electrical safety 600 V CAT III per IEC 61010-1, IEC 61010-
031, IEC 61010-032 and IEC 61010-2-033.
3.7. ELECTROMAGNETIC COMPATIBILITY
Emission and immunity in industrial environment according
to IEC 61326-1.
北京海洋兴业科技股份有限公司 电话:010-62176775