. The next time the
MEM
key is pressed, a disablement beep is emitted and
the
FULL
indication blinks.
4.8.5
READING THE STORED DATA
The data can be looked up using the
MR
function. See chapter 7.
4.9 MANAGEMENT OF THE ALARMS
The device has 3 distinct alarms that can be parameterized.
The (Ω, V, A) alarm thresholds are defined in the
SET-UP
menu, lines 4, 5, and
6; see §8.5. The alarms can be activated or deactivated in these same menus.
4.9.1
NO ALARM DETECTION
If no alarm is activated, the alarm icons
are not displayed.
When no alarm is triggered, the alarm
display displays the alarm threshold,
along with the direction of triggering (
, ) of the impedance, voltage, or
current alarm.
Fig. 8
4.9.2
VOLTAGE ALARM
If the voltage (product ZxI) exceeds the
threshold set, the alarm symbol and
the alarm threshold are displayed and
blink.
If the buzzer is active, the high-pitched
audible warning signal is emitted.
Fig. 9