40
Menu no. Indication Object and use
9
USE
Selecting the
Standard
or
Advanced
operating
mode
■
Enter the menu by ►.
USE
blinks.
■
Press ▲ or ▼ to select the
Standard
or
Advanced
mode.
-
Advanced
mode: the
icon is displayed.
-
Standard
mode
:
Std
is displayed.
■
Validate by ►. The return to the previous menu is
immediate.
10
FrEQ
Choice of transposition frequency for the
impedance in Advanced mode
■
Enter the menu by ►.
FrEQ
blinks.
■
Press ▲ or ▼ to select the transposition frequency
of the measured impedance from among the 4
possible values: 50, 60, 128, and 2,083 Hz.
■
Validate by ►. The return to the previous menu is
immediate.
11
HOLd
Activation/Deactivation of the
Pre-Hold
mode
■
Enter the menu by ►.
HOLd
blinks.
■
Press ▲ or ▼ to switch the
Pre-Hold
mode to
active or inactive.
-
Pre-hold
mode inactive: only the
icon is
displayed.
-
Pre-hold
mode active: the
and
icons
are displayed.
■
Validate by ►. The return to the previous menu is
immediate.
12
VER
Display of the version number
■
Enter the menu by ►.
■
The version number is displayed.
■
Return to the previous menu by ►.
Adjusting the clamp
The
SET-UP
menu enables users to adjust their clamps when they judge it
necessary. In order to avoid activating an adjustment procedure involuntarily,
there are 2 protections:
1. In the
SET-UP
menu, the
CAL
menu is placed last. A sustained press on ►
and ▲ and ▼ activates the
3 adjustment menus in
SET-UP
.
2. Once the adjustment menus have been activated, a long press (3 seconds) on
► is necessary to start each of the procedures.