According to the type of alarm selected, crossing of the alarm threshold will be
indicated by the appearance of      or       and a continuous beep at high or low fre-
quency.
Press the 

AL

 button again to exit the alarm function (the 

  

AL

  

 symbol disappears).

If the instrument is switched OFF, without the alarm mode being exited, this
configuration will be saved. The instrument will thus automatically switch the alarm
function ON again when the instrument is next switched ON.

Setting the alarm threshold

To programme the alarm threshold for resistance measurement it is necessary to first
press the 

ON + AL

 sequence.

The 

  AL

    symbol is displayed and the last digit of the threshold value flashes, as a

reminder that you are in threshold reset mode. This value is changed by pressing the

Ω 

Ω 

Ω 

Ω 

Ω 

button (increase) or 

A

 button (decrease). This new threshold value will be kept until

it is again changed, even after the clamp has been switched OFF.
The high alarm threshold can be between 1 and 199 

. It is programmed in at the

factory before delivery of the clamp.

OFF

 (instrument switched off) exits alarm threshold programming mode.

MEMORY

  

(

C.A 6415 exclusively

)

Zero reset of memory

Press the 

ON + MEM

 sequence for 3 seconds to display 

, this starts the clear

memoory procedure. At the 5th beep, the memory is cleared. Return to resistance
measurement is then automatic.

Recording a measurement

The recording of a measurement is only possible if the memory is switched on with the

ON + A

 sequence (

MEM 

symbol displayed).

Press 

MEM

 to record the displayed value. The number of values in memory goes up

by 1 and a long beep confirms the recording.

If the 

 symbol or the 

 symbol is continuously displayed, recording is

impossible because the clamp is not correctly closed or the battery is too low.

When the 99th value is recorded, the 

MEM

 symbol flashes to tell the user that the

memory is full. It is then no longer possible to record new values. If 

MEM

 is pressed

again there is no effect other than that a beep is emitted to indicate that recording is
prohibited. The memory should be reset to zero (after being re-read if necessary).

Summary of Contents for C.A 6410

Page 1: ...ANGE PINZA DI TERRA PINZA DE TIERRA C A 6410 C A 6412 C A 6415 F R A N C A I S E N G L I S H D E U T S C H I T A L I A N O E S P A N O L Notice de fonctionnement User s Manual Bedienungsanleitung Libr...

Page 2: ...leave the battery in the instrument in case of prolonged non use Opening of covers or removal of elements apart from those that can be manipulatedby hand risks giving access to parts that are dangero...

Page 3: ...15 only 35 Zero reset of memory 35 Recording a measurement 35 Reading recorded measurements 36 Special functions 36 Auto ON OFF 36 Buzzer ON OFF C A 6412 and C A 6415 36 Special symbols 37 SUMMARY OF...

Page 4: ...nditions of Sale available on request Reference GROUND TESTER C A 6410 P01 1220 11 GROUND TESTER C A 6412 P01 1220 12 GROUND TESTER C A 6415 P01 1220 13 Comes in carry case with a 9 V battery a set of...

Page 5: ...Measurement head 4 HOLD button Allows the user to freeze the displayed values and different functional symbols Used in combination with the ON OFF button secondary function allows the user to switch...

Page 6: ...o allows you to reset the memory to zero by holding the sequence ON MEM if buttons or A are pressed down and held the values scroll at a rate of 3 per second then 10 per second after they have been pr...

Page 7: ...u that the alarm is linked to the resistance measurement 24 set alarm threshold or alarm function mode 25 symbol showing that alarm threshold has been triggered by a low value 26 symbol showing that a...

Page 8: ...value I E Rloop Knowing E set by the generator and I measured we can deduce Rloop value displayed on the clamp More generally this principle makes it possible to find defective earths In fact the loo...

Page 9: ...measurement OL on the main display indicates that the value of resistance measured exceeds the measurement range on the main display indicates either that the clamp is open or that dirt is stopping th...

Page 10: ...d Low alarm indicates measurements below the alarm threshold High alarm indicates measurements above the alarm threshold When supplied the switch is set to the high alarm position To modify the type o...

Page 11: ...can be between 1 and 199 It is programmed in at the factory before delivery of the clamp OFF instrument switched off exits alarm threshold programming mode MEMORY C A 6415 exclusively Zero reset of m...

Page 12: ...ead memory mode SPECIAL FUNCTIONS Auto ON OFF Your clamp is fitted with a battery system for economising battery power After approximately 5 minutes operation of the battery if no button on the clamp...

Page 13: ...atically NOISE This symbol appears when a disturbing current which is too great circulates in the earth conductor i e when the product U Rloop I disturb Imeas is greater than approximately 40 V Resist...

Page 14: ...memory mode ON A Record measurement MEM MR Re read recorded measurements ON MEM press 2 s then or A Reset memory to zero ON MEM press 6 s Function saved even after the instrument is switched off Model...

Page 15: ...ore not guaranteed The memory is full because 99 values are recorded C A 6415 exclusively The alarm threshold set at 15 is on The low alarm arrow is displayed to indicate that this threshold has been...

Page 16: ...y carrying pylons the rails and sometimes even the fencing is connected to the earth In addition to get a very low value earth resistance an interconnection between pylons rails fencing is made thus c...

Page 17: ...uce a grounding system without equipotential defects To guarantee a perfect flow of loads to the earth the resistive values of the loops formed by these means must be low but also practically identica...

Page 18: ...60 Hz Case polycarbonate Dimensions 55 x 100 x 240 mm Max Clamping diameter 32 mm Weight approx 1 kg Watertightness IP30 Group III equipment as per EN 60529 Ed 92 IK04 as per EN 50102 Ed 95 Power sup...

Page 19: ...position Clamp horizontal Position of conductor in the clamp centred Proximity to metallic mass 10 cm Loop resistance Non choke resistance Current measured sinusoidal frequency 50 Hz Rate of distorsio...

Page 20: ...0 50 1 Indication of measurements up to approximately 0 07 but accuracy is not guaranteed below 0 1 Measurementrange 0 299 mA RMS 0 300 2 999 A RMS 3 00 29 99 A RMS Resolution 1 mA 1 mA 10 mA Accuracy...

Page 21: ...0 1 P R 50 60 Hz Frequency 47 800 Hz A 2 P 3 P R Disturbing current 3 L 5 L R 50 60 Hz Imeas Idisturb For I 1 A in the loop 2 x Rloop 40 V With R measured 30 Peak factor 3 1 4 to 2 5 A 1 P 1 5 P R 2...

Page 22: ...ed measurement with the value marked on the enclosed segment Proceed like this for each calibration value of the calibration loop Depending on the measurement differences read you can decide on the ne...

Page 23: ...you to check this instrument at least once a year For checking and calibration of your instrument please contact our accredited laboratories list on request or the Chauvin Arnoux subsidiary or Agent...

Page 24: ...ms Continuous 2 5 kHz 4 kHz 2 5 kHz 1 kHz 1 kHz 4 kHz Press button X Button prohibited Recording not possible X Noise Inter mittent Currentoverload Inter mittent Alarmcrossed low high alarm alarm Var...