5.
Use the arrow key
▲
to go to the Fixed Speed Mode Delay screen (image below). Press ENTER to
highlight the digit before ‘msec’.
6.
Use the arrow key
▲
or
▼
to change the value. Press ENTER when the desired value is shown; range
is 1 to 5000 milliseconds. The fixed speed mode delay time is now set.
IMPORTANT: A value other than ‘0’ in this field means the system will run in Fixed Delay Mode. When this value
is greater than 0, it is the delay (in milliseconds) between the time a bottle (or container) is detected to when the
UltraDoser
™
dosing system doses. The closer the container sensor is to the dosing head, the shorter the time delay.
7.
Use the arrow key
▲
or
▼
to go to the Dose Time screen (image below). Press ENTER to highlight the
value.
IMPORTANT: The Dose Time setting controls the amount of time (in milliseconds) that the UltraDoser system
dispenses liquid nitrogen. The amount of liquid nitrogen inside a container depends on many variables
including the position of the dosing head relative to the capper, the size of the dosing nozzle, the temperature of
the container contents, fill levels and head space. Therefore, the time setting for each filling operation must be
evaluated.