background image

Figure 3. Measuring surface 
resistivity.

Surface-to-Ground 
Measurements (RTG)

The Surface-to-Ground feature is 
best used when expected results 
are less than 10E6 ohms, and when 
residual voltages can be removed 
from the test surface. Residual 
voltages may cause the two upper 
range LEDs to light, resulting in a 
false reading.

When the Surface-to-Ground cord is 
plugged in, voltage is diverted from 
the center electrode to the cord in 
order to measure resistance. When 
making Surface-to-Ground tests, 
follow these procedures:

A. Check battery operation by 
depressing the white button. The red 
“On” lamp will indicate that the unit is 
functioning properly.

B. Plug the Surface-to-Ground 
cord into the left side of the 
megohmmeter.

Figure 4. Installing Surface-to-Ground 
cord assembly.

C. Place the Micro Meg on the test 
surface.

D. Connect the test cord to the 
ground-point for the surface.

E. Press the white button. The LEDs 
will indicate the test results. The 
measurement is in ohms.

Figure 5. Measuring 
Surface-to-Ground resistance.

F. Unplug the cord when testing is 
finished.

Battery Operation

Battery function is automatically 
checked by the unit when the meter 
is turned on. If the battery voltage 
drops below 8 VDC, the red LED on 
the far right of the meter will not light, 
indicating a “low battery” condition. 
Replace battery with a new 9V 
alkaline battery.

Maintenance

Your Micro Meg will require very 
little maintenance, and there are few 
user-serviceable parts. If your meter 
requires service beyond the simple 
procedures described below please 
contact the factory.

Electrode Cleaning

To assure accurate measurements 
the electrodes should be kept as 
clean as possible. A thorough 
cleaning at least every 3 months is 
recommended. The following is the 
recommended cleaning procedure:

1. Carefully remove the electrodes 
from their snap-jacks using a screw 
driver or other blunt flat bladed 
object. Avoid scratching or bending 
electrodes.

2. Clean both the snap stud and the 
electrodes with a cotton swab soaked 
in isopropyl alcohol. Check the rubber 
centers for wear - the rubber should 
form a convex spot on the bottom of 
the electrode. The three electrodes 
will be checked with each annual, 
factory calibration. If electrode 
replacement is recommended, 
the customer will be notified with 
estimated charges before any 
changes are made.

Theory of Operation

The Micro Meg measures resistance 
by setting up a bridge between two 
precision 1% resistors of known value 
and the surface tested, of unknown 
value. When the white button is 
depressed a test voltage of 9 VDC is 
applied to the megohmmeter’s center 
electrode or ground cord, depending 
on the test being performed. One of 
the bottom electrodes on the outside 
of the meter supplies the resistance 
bridge. The test voltage resulting 
from the bridge is compared to 
reference voltages established by 
six trimming potentiometers. Two 
integrated circuits are configured 
as comparators to check the 
voltage from the bridge against 
the potentiometers. Based on the 
setting of each potentiometer, the 
appropriate LED is lit.

Specifications

Sensitivity Range

Surface Resistivity
Less than 10E10 ohms/sq

Surface-to-Ground
Less than 10E9 ohms

LED Key (from right to left)

1. Red 
    >10E10, Meter On, and 
    Battery OK

2. Green 
    10E10 Static Dissipative

3. Green 
    10E9 Static Dissipative

4. Green 
    10E8 Static Dissipative

5. Green 
    10E7 Static Dissipative

6. Yellow 
    10E6 Static Dissipative

7. Yellow 
    10E5 Conductive

Snap Socket
Conductive Rubber
Snap Stud

Unit 17. Millbrook Business Park, Sybron Way • Crowborough, East Sussex TN6 3JZ United Kingdom

Phone: 00 44 (0) 1892-665313, Fax: 00 44 (0) 1892-668838 • Internet: 

Charleswater.co.uk

PPE-5046.E 

Page 2 of 3 

© 2011 CHARLESWATER LTD.

Figure 6. Cleaning the electrodes.

Reviews: