PPE-5034.E
Page 2 of 2
© 2015 Charleswater
Unit 17. Millbrook Business Park, Sybron Way • Crowborough, East Sussex TN6 3JZ United Kingdom
Phone: 00 44 (0) 1892-665313 • Internet:
Testing the Footwear Circuit
1.
Plug one of the included test leads into the black
banana jack labeled “E” on the Calibration Unit.
Connect the opposite end of the test lead into the
yellow banana jack labeled “FOOT PLATE” on the
Combo Tester. Toggle the switch to the “FOOT
PLATE” position.
Specifications
Accuracy:
±5%
Dimensions:
8.1 cm x 11.2 cm x 2.3 cm
Weight:
.1 kg
Figure 3. Verifying the calibration of the wrist strap
circuit in the Combo Tester
Figure 4. Connecting the test lead from banana jack “E”
on the Calibration Unit to the FOOT PLATE banana jack
on the Combo Tester
2.
Connect the second test lead to the red 675K
banana jack labeled “A” on the Calibration Unit.
Touch the opposite end of the test lead to test plate
on the Combo Tester and press down to activate
the test. Be sure not to touch the banana plug or
test plate with your skin.
3. Observe the response from the Combo Tester. The
alarm should sound, and the LOW LED should
illuminate red.
4.
Repeat the procedure for test points B (825K), C
(35M) and D (40M). The expected test results can
be found in the following table.
Test
Point
Resistance
Value
Test Output
A
675K
FAIL LOW
B
825K
PASS
C
35M
PASS
D
40M
FAIL HIGH
Figure 5. Verifying the calibration of the footwear circuit
in the Combo Tester
Limited Warranty, Warranty Exclusions, Limit of
Liability and RMA Request Instructions