R460 Keg Scale
.
10
SETTINGS CONFIGURATION
1. Changing the default measuring unit (kg & lb). (Default: kg)
While scale is on, press
TARE
key for 6 seconds to enter setting.
At this time, display shows “SET”, followed up by “UNIT”.
a. Press
TARE
to enter setting mode “UNIT”.
b. Press
HOLD
to switch between weighing unit:
kg, lb, lb oz
.
c. Press
TARE
to select unit.
d. Press
HOLD
to turn chosen unit
On
or
Off
to be default weighing unit.
e. Press
TARE
to select.
f. Press
HOLD
to cycle back to “UNIT”.
g. Press
HOLD
7 times to “END”, then press
TARE
to save setting and
restart the scale.
2. Programming the Auto Off Time. (Default: 180S)
While scale is on, press
TARE
key for 6 seconds to enter setting.
a. Press
HOLD
to cycle to mode “AOFF”.
b. Press
TARE
to enter setting.
c. Press
HOLD
to view options:
120S, 180S, 240S, 300S, OFF
.
d. Press
TARE
to select.
e. Press
HOLD
6 times to “END”, then press
TARE
to save setting and
restart the scale.
3. Toggling the Beep Sound.
(Default: ON)
While scale is on, press
TARE
key for 6 seconds to enter setting.
a. Press
HOLD
to cycle to mode “bUrr”.
b. Press
TARE
to enter setting.
c. Press
HOLD
to choose
On
or
Off
.
d. Press
TARE
to select.
e. Press
HOLD
5 times to “END”, then press
TARE
to save setting and
restart the scale.
3. Programming the Display Backlight.
(Default: AUTO)
While scale is on, press
TARE
key for 6 seconds to enter setting.
a. Press
HOLD
to cycle to mode “bAKL”
.
b. Press
TARE
to enter setting.
c. Press
HOLD
to view options:
Auto, On, OFF
.
d. Press
TARE
to select.
e. Press
HOLD
4 times to “END”, then press
TARE
to save setting and
restart the scale.
[PSET] [t_lr] [CALb3]
SETTINGS ARE FOR DISTRIBUTOR AND SERVICE PERSONNEL USE ONLY.