Charder MS3450 User Manual Download Page 9

 

-31- 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

MEASURING UNDER 121 cm 

1. Fold the plate and press 
the buckle. 

2. Pull down head piece 
while buckle is pressed 

3.  Height reading can be 

taken at 

point.

 

 

 

-8-

Guidance and manufacturer’s declaration-electromagnetic immunity

 

The MEDICAL SCALE MS3450 is intended for use in the electromagnetic 
environment specified below.   
The customer or the user of the MEDICAL SCALE MS3450 should assure that it is 
used in such an environment.   

Immunity test

IEC 60601 test 

level

 

Compliance 

level

 

Electromagnetic 

environment-guidance

 

Electrostatic 
discharge(ESD) 
IEC 61000-4-2 

±

  8 kV contact 

±

 2 kV, 

±

 4 kV, 

±

 8 kV, 

±

 15 kV 

air 

± 6 kV contact   
± 8 kV air 

Floors should be wood, 
concrete or ceramic tile. 
If floors are covered with 
synthetic material, the 
relative humidity should 
be at least 30%   

Electrical fast 
transient/burst 
IEC 61000-4-4 

± 2kV for power 
supply lines + 
1kV for 
input/output 
lines  

± 2kV for power 
supply lines Not 
applicable  

Mains power quality 
should be that of a typical 
commercial or hospital 
environment.  

Surge IEC 
61000-4-5  

± 1kV line(s) to 
line(s) ± 2kV 
line(s) to earth   

± 1kV differential 
mode Not 
applicable  

Mains power quality 
should be that of a typical 
commercial or hospital 
environment.  

Voltage Dips, 
short 
interruptions 
and voltage 
variations on 
power supply 
input lines IEC 
61000-4-11  

0% UT for 0,5 
cycle 
0% UT for 1 
cycle 
 
70% UT(30% 
dip in UT) for 25 
cycles 
 
0% UT for 5 s   

<5% UT(>95% 
dip in UT) for 0,5 
cycle 40% 
UT(60% dip in 
UT) for 5 cycles 
70% UT(30% 
dip in UT) for 25 
cycles <5% 
UT(>95% dip in 
UT) for 5 s   

Mains power quality 
should be that of a typical 
commercial or hospital 
environment. If the user 
of the MEDICAL SCALE 
MS3450  requires 
continued operation 
during power mains 
interruptions, it is 
recommended that the 
MEDICAL SCALE 
MS3450 be powered 
from an uninterruptible 
power supply or a 
battery. 

Summary of Contents for MS3450

Page 1: ...amended by 2007 47 EC Medical Device Directive Please see separate document showing on sticker of device for above CE marking Authorized EU Representative Manufactured by Charder Electronic Co Ltd No...

Page 2: ...C guidance and manufacturer s declaration 7 SPECIFICATIONS 12 PANEL 13 KEY FUNCTIONS 13 LCD DISPLAY SYMBOLS 14 SYMBOLS information 15 OPERATION 15 SETTING UP 20 PRINT function and data transmission 21...

Page 3: ...hing value can be read off after a stable weighing value has been obtained Before any use the scale must be checked for correct condition by the authorized person SAFETY INSTRUCTION Before putting the...

Page 4: ...m the mains power supply before cleaning 36 3 Connection failure for data transmission to PC or printer Wrong connection wires or faulty wires for transmission between the digital indicator load cells...

Page 5: ...her external objects within weighing area The weighing scale is not put in a solid firm ground area such as carpet floor or lawn 4 MAINTENANCE The scale does not require any routine maintenance Howeve...

Page 6: ...for operation 2 If the problem still persists inspect soldering of controller PCB or replace the controller PCB Overload This indicates that the scale s load sensor s have been overloaded Reduce the...

Page 7: ...pplied at the device Number as an example Please note the accompanying documents or Observe operating instructions Charder Electronic Co Ltd No 103 Guozhong Rd Dali Dist Taichung City 412 Taiwan R O C...

Page 8: ...ment RF emissions CISPR 11 Class B The MEDICAL SCALE MS3450 is suitable for use in all establishments including domestic establishments and those directly connected to the public low voltage power sup...

Page 9: ...burst IEC 61000 4 4 2kV for power supply lines 1kV for input output lines 2kV for power supply lines Not applicable Mains power quality should be that of a typical commercial or hospital environment S...

Page 10: ...ironment guidance Conducted RF IEC 61000 4 6 3 Vrms 150 KHz to 80 MHz 6 V in ISM bands between 0 15 MHz and 80 MHz 80 AM at 1 kHz 3 Vrms Portable and mobile RF communications equipment should be used...

Page 11: ...Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicte...

Page 12: ...er W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 7 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10...

Page 13: ...Capacity 300 kg x 0 1 kg Accuracy 200g Weight Unit Kg lb LCD Display 1 4 LCD digital display Dimension Platform size 310 x 310 mm Column Height 900mm Key Functions UNIT SEND ON OFF ZERO HOLD BMI TARE...

Page 14: ...area of human body For many clinical purpose BSA is the better indicator of metabolic mass than body weight because it is less affected by abnormal adipose mass BSA is commonly used in medicine parti...

Page 15: ...daptor AC adapter 1 6 User manual IN 00011 1 14 UNIT Key to switch kg or lb and to change height value under BMI mode HOLD BMI HOLD key and BMI Body Mass Index key LCD DISPLAY SYMBOLS BSA Body surface...

Page 16: ...l of battery power Stability sign Negative sign Zero indication Battery symbol 24 8 Printout Data Through USB cable connector Click OK button to complete the setting The PC will retrieve the weighing...

Page 17: ...reading can easily be read Note Determination of average value is not possible when a person moves too much HOLD function won t work if the weight is under 2 kg 1 Switch on the scale The diagnostic se...

Page 18: ...nds First display will show HOLD and lock weight then display show height value 4 Adjust height value by using and keys 5 Switch between digits by using key 6 Press to confirm height value 7 Display w...

Page 19: ...C printer com port with USB cable connector transmission For use on Windows Vista or higher please download hyper terminal software program from Charder website http www chardermedical com download dl...

Page 20: ...2 Press key and hold for 6 seconds to enable setting mode 3 LCD shows SETUP first and then A OFF appears Switch between modes and menu Reverse switch between modes and menu Confirm setting MENU AUTO O...

Reviews: