IV. Using Device
A. Basic Operation
Switch on the device using
[ON/ZERO/OFF]
key. The device will
automatically perform self-calibration, displaying software version.
Once "0.00 kg" appears on indicator, device is ready for measurement.
Note
: If "0.00 kg" does not display on indicator, press
[ON/ZERO/OFF]
key to zero the device. This function can be used for weight within
±
2%
of full capacity.
Carefully place subject upon the measurement platform. After the weight
has stabilized, the "stable" symbol will appear on indicator.
Note
: If subject's weight exceeds scale capacity (including tare),
indicator will display "Err" prompt due to overload.
B. Tare
The tare function allows the user to deduct the weight of objects from the
device's measurement result. Tare can be used when weight of object is
≧
at/above 2% of 20 kg capacity.
1. Place object that needs to be tared onto measurement platform.
2. Press
[ON/ZERO/OFF]
key after stable symbol appears on indicator.
Display will indicate "0.00 kg".
3. Place subject (plus tared object) to be weighed upon measurement
platform. Conduct measurement.
4. To clear tare value, remove all objects from measurement platform,
and press
[ON/ZERO/OFF]
key.
C. Hold
The hold function determines average weight, designed to be used if
subject's weight will not stabilize (ex: an active infant).
Note:
if fluctuation is too severe, average weight determination will be
difficult and hold may not function correctly
1. Switch on the device normally.
2. Press the
[HOLD]
key. "HOLD" will be displayed on the indicator.
3. Carefully place subject on measurement platform.
4. After a few seconds, the average weight will be displayed on the
indicator. This weight will be locked - at this point, subject can be