8
Guidance and manufacturer’s declaration-electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The
customer or the user of the device should assure that is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic
environment-guidance
Radiated RF IEC
61000-4-3
3 V/m
80MHz to 2,7 GHz
3 V/m
80MHz to 2,7
GHz
Recommended separation
distance:
d = 1,2 √
P
d = 1,2 √
P
80MHz to 800 MHz
d = 2,3 √
P
800MHz to 2,5 GHz
Where
P
is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and
d
is the recommended separation
distance in metres (m).
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey
a
,
should be less than the
compliance level in each
frequency range
b
.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.