C h a p t e r   6    

:   M a i n t e n a n c e  

1. 

The storage of pH Glass Electrode:  

 

 

 

 

 

 

  Short-term  : Stored in the pH = 4 buffer solution;  

 

 

 

  Long-term  : Stored in the pH = 7 buffer solution.  

2. 

pH Glass Electrode Cleaning : 

 

 

 

 

 

 

 

  pH glass electrode cleaning glass electrode bulb contamination may make the electro
  de response time longer. CCl4 or soap can be used to wipe the dirt, and then immer
  sed in distilled water a day and night to continue to use. When the pollution is serious, 
  can be 5% HF solution for 10 to 20 minutes, immediately rinse with water, and then 
  immersed in 0.1N HCl solution for a day and night to continue to use.  

3. 

Glass Electrode Aging Treatment : 

 

 

 

 

 

 

  The aging of the glass electrode and the gradual change in the structure of the glue 
layer. Old electrode response is slow, film resistance is high, slope is low. Exfoliation of 
the outer layer with hydrofluoric acid can often improve electrode performance. If this 
method can be used to regularly remove the inner and outer layers, the electrode life is 
almost unlimited. 

4. 

The Storage of the Reference Electrode:  

 

 

 

 

 

  Silver - silver chloride electrode The best storage solution is saturated potassium chlo-
ride solution, high concentration of potassium chloride solution can prevent the silver 
chloride in the liquid junction precipitation, and maintain the liquid junction in the work 
status. This method also applies to the storage of composite electrodes. 

5. 

The reference electrode regeneration reference electrode problems caused by the vast 
majority of liquid junction caused by blockage, the following methods can be resolved: 

 

Soaking Fluid Interface: 

10% saturated potassium chloride solution and 90% distil-

led water mixture, heated to 60 ~ 70 °C, the electrode immersed in about 5cm, soak 
for 20 minutes to 1 hour. This method dissolves the crystallization of the electrode tip. 

 

Ammonia Soaking:

 When the liquid interface is blocked by silver chloride can be leac-

hing with concentrated ammonia. The specific method is to clean the electrode, the 
liquid vent after immersion in ammonia 10 to 20 minutes, but do not let ammonia into 
the electrode inside. Remove the electrode with distilled water to wash, re-add the in-
ternal liquid and continue to use. 

 

Vacuum Method:

 the hose to match the reference electrode fluid interface, the use of 

water suction pump, suction part of the liquid through the fluid interface, remove the 
mechanical blockage. 

 

Boiling Fluid Junction: 

silver - silver chloride reference electrode liquid interface im-

mersed in boiling water for 10 to 20 seconds. Note that the next time you boil, the 
electrode should be cooled to room temperature. 

 

When the above methods are invalid, sandpaper grinding can be used to remove the 
mechanical method of grinding. This method may cause the sand under the grinding 
into the liquid interface. Causing permanent clogging.

 

Summary of Contents for pHI01

Page 1: ...CYK Tech Online pH ORP Meter User s Manual pHI01 Copyright CeYeKo Technology 2021 pHI01_V0 0 Committed to Process Automation Solutions...

Page 2: ...se operation Note Modification of this manual s contents will not be notified as a result of some factors such as function upgrading We try our best to guarantee that the manual content is accurate if...

Page 3: ...kes If the instrument is operated in other ways not described in the manual the protections that the instru ment give may be destroyed and the failures and accidents incurred due to violation of preca...

Page 4: ...tric shock Cut off electric powers before making any checks to avoid electric shock Check the condition of the terminal screws regularly If it is loose please tighten it be fore use It is not allowed...

Page 5: ...t and controlled equipment Each part of the instrument has a certain lifetime which must be maintained and repai red on a regular basis for long time use The product shall be scrapped as industrial wa...

Page 6: ...loss of parts or unpre dictable damage caused directly or indirectly by improper operation of the user Discla imer After opening the box please confirm the package contents before starting the operati...

Page 7: ...allation Methods 11 2 2 Electrode Installation 12 2 3 Wiring 13 2 3 1 Identification of Terminals 13 Chapter 3 Navigation Keys 14 3 2 Definition of Buttons 14 Chapter 4 System Menu Operating 15 4 1 Mo...

Page 8: ...the RS485 interface through the MODBUS RTU protocol can be easily connected to the computer to achieve monitoring and recording Chapter 1 Introduction After opening the box please confirm the package...

Page 9: ...tional Accuracy pH 0 02pH ORP 1mV Input Resistance 10 Temperature Compen sation NTC10K 10 60 C Accuracy 0 3 C NTC10K 60 130 C Accuracy 2 C PT1000 Accuracy 0 3 C 10 130 C Manual Automatic Output 4 20mA...

Page 10: ...this product from rising please install it in a well ventilated place When installing this product please do not tilt it to the left and right try to install it ho rizontally it can be tilted back 30...

Page 11: ...CYK Tech www ceyeko com Page 11 2 1 3 Installation Methods Install a 92 5 92 5 mm mounting hole on the instrument cabinet or mounting panel The thickness of the mounting panel is 1 5mm 13mm...

Page 12: ...n Schematic Diagram of Common Installation Method Side wall installation Flange mounted at the top Pipe installation Top installation Submersible installation Flow through installation The interface...

Page 13: ...C10K and PT1000 connect here C Temperature compensation terminal C PT1000 three wire temperature gro unding PT1000 two wire need to be short connected to TEMPB not NTC10K NC Unidentified 485A RS485 co...

Page 14: ...Return to the previous page Under Calibration Page Skip this item RIGHT Make a recurrent selection of digit of parameters modify the original indication value MENU Under Monitoring Page Enter the MEN...

Page 15: ...instrument is equipped with monochrome lattice LCD 128 64 resolution Push MENU to enter password verification page input password to enter the home page Push EXIT to enter alarm inquiry page to inqui...

Page 16: ...Please contact us if you forget your password System Setting Settings of language buzzer and backlight modification of password and factory settings Signal Setting Toggle electrode type and automatic...

Page 17: ...nd log in with new password Factory Setting Return to factory settings 4 3 3 Signal Set ting Electrode Type Set the electrode type pH electrode and ORP electrode two types Temperature Compensation Set...

Page 18: ...ion Enter the ORP calibration screen the first ORP electrode into the 86mV standard solution standing for a moment to be shown after the stability press the ENT then the ORP electrode into the 256mV s...

Page 19: ...ed value is greater than the low reported off value the low alarm relay off ORP High Alarm When the measured value is greater than the high reported pull value the high reported relay pull when the me...

Page 20: ...s Function Code Register Address Number of Data CRC Check Data ADDR 0x03 M N CRC 16 Number of Bytes 1 1 2 2 2 Return Format Definition Address Function Code Register Address Number of Data CRC Check D...

Page 21: ...e CRC check value Example of Temperature Reading Computer Sends 00 03 00 02 00 01 24 1B Conductivity Meter Returns 00 03 02 00 FA 05 C7 Return Command Annotation 00 is the address of slave which can b...

Page 22: ...uid junction in the work status This method also applies to the storage of composite electrodes 5 The reference electrode regeneration reference electrode problems caused by the vast majority of liqui...

Page 23: ...measures Conductivity instrument can not be calibrated Answer The standard solution is not mixed in a correct way or the electrode is damaged The instrument can not measure accurately after calibratio...

Page 24: ...CYK Tech CeYeKo Fluid Measurement Technology www ceyeko com Copyright CeYeKo Technology 2021 Online pH ORP Meter Manual pHI01_V0 0...

Reviews: