Measures for Safe Operation
2.4.5.1
Electromagnetic Emissions
The
NE
MESYS XL module may be used in any type of facility, including living
quarters, as well as those that are connected to the public utilities network that
also supplies buildings used for living purposes.
2.4.5.2
Electrostatic Discharge
Flooring materials should be wood, concrete or ceramic tiles. If the flooring is
made of a synthetic material, the relative humidity must be at least 30%.
2.4.5.3
Electric Disturbances
The quality of the supplied voltage should conform to the standard of a typical
business or hospital environment.
2.4.5.4
Magnetic Disturbances
Do not place power connector cables, including those of other appliances, in
close proximity of the device and its cables. Portable and mobile communication
devices may not be used in closer proximity of the unit or its cables than the
recommended safety distance for said devices.
2.4.6.
Device Safety Features
In case of an emergency, the device may be switched off at any time using the
mains switch (rocker switch) on the back of the housing or by pulling the power
supply connector from the device. This will cause no damage to the device itself.
Additionally, the device features a safety cover that shields all moving parts and
reduces the risk of injury. Removing the cover will stop and switch off the device.
Using the pressure sensor interface, the device may be programmed to
automatically stop if a preset maximum pressure is exceeded.
The device has a built-in overload clutch that will disconnect the driving unit and
switch off the device.