Ceterix NovoStitch Instructions For Use Manual Download Page 4

 

 

c.

 

While maintaining light proximal tension on the loop, slightly squeeze Lever B so that 

the needle moves forward about 1 mm within the lower jaw (see Figure 4), then release 

Lever B.  Do not fully advance Lever B at this stage. The suture strand will be locked in 

the lower jaw by the needle. 

 

Figure 4:  Slightly squeeze Lever B to advance needle approximately 1mm. 

 

 

d.

 

Retract the lower jaw by pushing down on Lever C (see Figure 1), the lower jaw 

retractor.   

e.

 

The suture is now loaded for passing suture through soft tissue. 

 

3.

 

With the suture loaded and the lower jaw in the retracted position, pass the device into the surgical 

site under arthroscopic visualization. Use Lever A to position the upper jaw parallel to the shaft 

during insertion of the device. 

 

B.  DIRECTIONS FOR USE PASSING SUTURE THROUGH MENISCAL TISSUE

 

 

1.

 

Ensure portal placement and preparation is appropriate to access the region of the meniscus to be 

sutured. 

 

2.

 

Advance the upper jaw along the superior surface of the meniscus following the radius of the 

femoral condyle. Release Lever A as the upper jaw is positioned between the superior surface of the 

meniscus and the femoral condyle. 

 

3.

 

When the upper jaw is in position, with all fingers on the white handle (not on Lever A) fully advance 

the lower jaw by squeezing Lever B (see Figure 1 above). Ensure that the lower jaw slides 

underneath the meniscus, then release Lever B. 

 

4.

 

When both upper and lower jaws are in position, simultaneously squeeze Lever A and apply slight 

forward pressure to prevent the tissue from migrating distally. Arthroscopically observe upper jaw 

movement. Lack of upper jaw compression during upper jaw clamping is a sign that the tissue is too 

P/N:  LBL-0012     Rev: E 

 

 

Page  4 

 

Reviews: