Cerlic pHX Manual Download Page 15

 

                                                                                                         

 

- 15 - 

13.

 

Calibration 

The sensor is shipped pre-calibrated, but since all electrodes are different, the sensor 

must be calibrated with its electrode to give a good measurement. The electrode 

degrades by time and calibrations shall be done regularly following a maintenance 

schedule. The time between calibrations varies with the applications, it is usually 

between one and ten weeks. It is important that the sensor has been switched on for 

about 30 minutes prior to the calibration in order to stabilize. 

Calibrating pH 

For a pH calibration there are two alternatives: smart or manual. Irrespective of the type 

of calibration you chose, you need two pH buffers which both have a common, known 

temperature. 
For a smart calibration you need two different buffer solutions with pH 4.00, 7.00, or 

10.00. 
For a manual calibration, two different buffer solutions can be used irrespective of the 

buffer value. 
Remove the sensor and, if necessary the electrode as well, but make sure that the 

electrode is still connected to the signal amplifier. Make sure that you have at least 2 

standard buffers available with pH 4.00, 7.00 or 10.00 of a known temperature (it is best 

to pour the buffers into plastic cups). 
 

 

Enter the sensor menu. The output is frozen when the menu is opened. 

 

Open the “Calibrate” menu. 

 

Select "Calibrate", then "Smart" or ”Manual” and press ENTER. 

 

A window pops up and asks for the buffer temperature. Enter the temperature and 

press ENTER 

 

BB2 will say ”Put the electrode in buffer 1”. Spray the electrode using clean water, 

then immerse it in the first buffer, and press ENTER.  

 

The text will say "Waiting for stable signal xxx,x mV", xxx,x is the millivolt signal 

from the electrode. By pressing ENTER you can select to abort the calibration or to 

accept the value shown.  

 

If smart calibration is used and BB2 recognizes the buffer, BB2 will suggest a pH 

value showing the text ”Accept smart calibration pH X.X”. You can choose to 

accept the value or to change it manually. 

 

If manual calibration is used BB2 will always ask for the value of the buffer, enter 

the value and press ENTER. 

 

BB2 will now tell you ”Put the electrode in buffer 2”. Spray the electrode using 

clean water, then immerse it in the second buffer, and press ENTER. 

 

The text will say "Waiting for stable signal xxx,x mV", xxx,x is the millivolt signal 

from the electrode. By pressing ENTER you can select to abort the calibration or to 

accept the value shown.  

Summary of Contents for pHX

Page 1: ...2019 02 20 C60C5EN19 pHX ReX pH Redox Sensor...

Page 2: ...e for cleaning solenoid valves 8 10 Maintenance 9 Error messages 9 Broken Electrode 9 Incorrect Measurements 9 11 Sensor display 10 12 Menu for Sensors 11 Menu for pHX 11 Menu for ReX 13 13 Calibratio...

Page 3: ...de which in combination provide a voltage in linear proportion to the pH value Electrodes internally combined with a reference and measuring cell are divided into two categories gel electrodes and ref...

Page 4: ...05264 Flushing system 1 meter excl solenoid valve P N 11205612 Flushing system 2 meter excl solenoid valve P N 11205903 Solenoid valve 220V for flushing P N 11705516A Incoming hose 16mm 5 8 to sensor...

Page 5: ...ations where the level varies greatly a mounting kit with a float ball should be used Take care to mount the mounting kit in such a manner that it is easily accessible in conjunction with calibration...

Page 6: ...s may need to be rinsed away with water If the electrode has holes for refilling the electrolyte inside the electrode should be checked and refilled The level should be at least 1 cm 1 2 below the ref...

Page 7: ...dia Then tighten the nut with your hand If you have an electrode which does not have a hard wired cable the protective cover on top of the electrode should be removed and the enclosed plug in cable sh...

Page 8: ...being the master If needed specify the extra freeze time in seconds If the sensor shall be cleaned as master specify the relay to be used according to the wiring inside the control box For example if...

Page 9: ...calibration the pHX does not perceive a difference of buffers Make sure that you use two different buffers during calibration The electrode may be broken The electronic components in the sensor may b...

Page 10: ...has two information pages the first one shows in addition to the measured value Max Min and average values for the last 24 hours The values are based on hour values which means they may jump when a ne...

Page 11: ...he first buffer Sample 2 pH Actual pH for the second buffer Calibrated Date of last air calibration New eletcr Date when electrode was last changed Push ENTER to edit the date then ENTER again to stor...

Page 12: ...eaning System Type Type of sensor read only Serial Serial number of the sensor read only SoftW Software version of the sensor read only Info Press ENTER to go to info read only menu This menu is for C...

Page 13: ...eaning cycles Length sec 0 999 seconds duration of flushing cycle Freeze sec 0 999 seconds extra freeze time of output signal after a flushing cycle Relay 1 2 Along 1 or Along 2 Select relay to operat...

Page 14: ...are version of the sensor read only Info Press ENTER to go to info read only menu This menu is for Cerlic internal use RV 0 mV Raw value for 0 mV RV 1000 mV Raw value for 1000 mV Ch1 Raw value ORP mV...

Page 15: ...4 00 7 00 or 10 00 of a known temperature it is best to pour the buffers into plastic cups Enter the sensor menu The output is frozen when the menu is opened Open the Calibrate menu Select Calibrate t...

Page 16: ...se you need a buffer with a known potential Remove the sensor and if necessary the electrode as well but make sure that the electrode is still connected to the signal amplifier Wipe clean the electrod...

Page 17: ...ciple Electrode 12mm Cable connection 5 pin M12 plug Cable length 10 m 33 ft Cable material Polyurethane Enclosure IP65 ReX P N 11305598 Tube and electrode holder SIS2 343316SS Top PVC Dimensions 28 x...

Page 18: ...3 68 EEC 89 336 EEC EMC Directive as amended by 92 31 EEC and 93 68 EEC The following standards and or technical specifications have been applied EN 61000 6 4 2001 Electromagnetic compatibility EMC Pa...

Page 19: ...19 16 Dimensions...

Page 20: ...20 Appendix 1 Optional slide rail holder...

Page 21: ...roxide ions the output from the electrode is 0 0 mV and pH is 7 The millivolt signal is measured by pHX and the corresponding pH is calculated in BB2 mV pH H ions Mol l OH ions Mol l Example 414 0 1 0...

Page 22: ...3 1 70 2 76 3 82 4 88 5 94 7 00 7 94 8 88 9 82 10 76 11 70 12 64 55 131 1 55 2 64 3 73 4 82 5 91 7 00 7 91 8 82 9 73 10 64 11 55 12 46 65 149 1 40 2 52 3 64 4 76 5 88 7 00 7 88 8 76 9 64 10 52 11 40 1...

Page 23: ...n the main display Go to the sensor menu it is accessed by pressing ENTER for five seconds Select System and press ENTER Type _____________________________________________________ Serial _____________...

Page 24: ...aning sub menu of the sensor menu the following parameters can be set Cleaner _____________________________________________________ Cleaning interval __________________________________________________...

Reviews: