3
Important Safeguards
Important Safeguards
Microwave sample preparation imposes a unique set of safety considerations beyond the basics of good laboratory
practice. Additional Safeguards for the MARS 6 instrument are included in the manual, PN 600832. General
guidelines for safe operation of the iPrep are presented below and all specific safety messages are located
throughout the manual.
Proper precautions must be taken to avoid contact with reagents or reagent vapors. Protective gear should be worn
as outlined in the user’s safety program for hazardous materials and the reagent manufacturer’s safety data sheet.
Refer to these guidelines for proper handling and disposal of the reagents.
Operating Parameters
iPrep 12
iPrep 16
iPrep Parameters
Minimum Volume
5 mL
5 mL
Maximum Volume
30 mL
30 mL
Maximum Pressure
1500 psi
1500 psi
Maximum Hold Time
45 min
45 min
Maximum Temperature
Maximum Programmable Temperature
300
º
C
300
º
C
Nitric Acid
250
º
C
250
º
C
Hydrochloric Acid*
250
º
C
250
º
C
Hydrofluoric Acid*
250
º
C
250
º
C
Sulfuric Acid
300
º
C
300
º
C
Phosphoric Acid
300
º
C
300
º
C
Aqua Regia*
250
º
C
230
º
C
Water
180
º
C
180
º
C
Maximum Sample Weight
Food, Plant, and Animal Tissue
1.5 g
1.5 g
Oil and Plastic
0.5 g
0.5 g
Pharmaceutical
1.5 g
1.5 g
Inorganic
1.5 g
1.5 g
Metallurgical
1.5 g
1.5 g
Clinical
1.5 g
1.5 g
* The maximum achievable temperature is 210 °C if running more than 6 vessels.
Sample Preparation Safeguards
• Do not use water to dilute the acid(s) greater than 25% when running over 220 ºC or to add volume before
digestion. This can cause uneven heating of vessels
• Run similar sample types with the same acid(s) and acid volume in each vessel
• Optimal Sample Volume is 10mL and will result in more even heating
• Minimum acid volume is 5 mL
• Maximum volume is 30 mL
• Maximum sample weight varies per sample type. See "Operating Parameters" for weights
• Acid blanks cannot account for more than 25% of turntable when mixed with samples
• Acid blanks cannot be run with organic samples > 230 °C
• Acid blanks cannot be run with samples when acid mixtures are > 50% hydrofluoric acid. Acid blanks must
be run separately