celesco PT1 Series Replacement Instructions Download Page 2

Page 2 of 4

MEASURING CABLE REPLACEMENT

1.

 

Remove the two socket head cap screws from the cover and set aside.

2.

 

Gently lift off the cover.  There are still wires connected between the rotational
sensor in the unit and the cover; be careful not to force or pull too hard.  If the
cover is stuck use a twisting motion.

3.

 

If the cable is still intact and can be extended and retracted, proceed with step 4.
Otherwise, skip to step 32.

4.

 

There are two set screws offset 90 degrees holding the spool to the potentiometer
shaft.  While looking through the access hole in the frame, extend the cable so one
of the set screws can be seen.

5.

 

With an allen wrench, loosen the set screw.

6.

 

Extend the cable again until the other set screw is visible and loosen the set screw

7.

 

Grasp the potentiometer and its mounting plate and pull straight out of the spool.
The screws and spacers on either end of the mounting bracket don’t have to be
loosened or removed.

8.

 

Check potentiometer for damage.  If cable has been free-released or broken, the
potentiometer may be broken.  Turn the potentiometer shaft all the way in both
directions.  If the transducer is a 2 inch or 5 inch range, the shaft should rotate
smoothly with no resistance.  If the transducer is a 10 inch or greater range, the
shaft should rotate smoothly with no resistance until it suddenly stops.  If rough
spots, resistance, or “soft” stops are felt, the potentiometer is damaged and must
be replaced.  Contact Celesco for new potentiometers and replace if necessary.

9.

 

Extend the measurement cable a short distance and cut it with a pair of wire
cutters.

CAUTION

The spring motor can be dangerous while it has tension.  The spring, if allowed to, can
suddenly uncoil itself by shooting out of the spring housing.  It’s sharp and can cause
injuries.  Be careful when doing anything with the spring exposed.  Wear gloves and
safety goggles.

10.

 

Loosen the #8-32 screws holding the spring housing. If the spring still has
tension, when the bolts are removed the spring housing will spin.  Get a firm grip
on the frame and spring housing remove the bolts and set aside.

11.

 

Carefully let the spring housing rotate until all spring tension is released, while
keeping the housing against the spring side plate.

12.

 

The spring is probably still hooked to the spring arbor.  BE CAREFUL TO NOT
pull the spring housing off yet or the spring might shoot out of the housing.

13.

 

Separate the spring housing from the frame just enough to get a flathead
screwdriver between the housing and the spring side plate.

14.

 

Insert the flathead screwdriver and use it to slide the spring off the spring arbor
and into the spring housing.

15.

 

Remove the spring/spring housing and set aside.

16.

 

Remove the spring side plate and set aside.

Summary of Contents for PT1 Series

Page 1: ...EASURING CABLE REPLACEMENT INSTRUCTIONS FOR PT1 SERIES For Models PT1A PT1DC PT1MA Celesco Transducer Products 7800 Deering Avenue Canoga Park CA 91309 Tel 800 423 5483 Fax 818 340 1175 Created 4 19 00 C data PT1man ...

Page 2: ...haft should rotate smoothly with no resistance until it suddenly stops If rough spots resistance or soft stops are felt the potentiometer is damaged and must be replaced Contact Celesco for new potentiometers and replace if necessary 9 Extend the measurement cable a short distance and cut it with a pair of wire cutters CAUTION The spring motor can be dangerous while it has tension The spring if al...

Page 3: ...e 25 Remove the spring side plate and set aside 26 While looking through the access hole in the frame rotate the spring arbor with fingers until one of the set screws is visible 27 Using an allen wrench loosen the set screw 28 Rotate the arbor again until the other set screw is visible 29 Loosen the set screw 30 Pull the potentiometer straight out of the spool 31 Check potentiometer for damage If ...

Page 4: ... not to over tighten or the plastic spring housing may be damaged 48 Fully extend and retract the cable a few times to ensure smooth cable travel 49 Loosen the large nut holding the potentiometer to the pot mounting plate just enough so it can rotate freely 50 If spacers and 4 40 screws that hold the mounting bracket in place were removed replace them now 51 Replace the potentiometer and mounting ...

Reviews: