p.4

3.2 MECHANICAL INSTALLATION

Install the SLR Pressure Transducer Display as follows:

1.   Remove the display from the box. Remove the outer protective bag and 

discard.

2.   Follow standard cleanroom entry procedures where applicable.

3.   Install the display on the transducer. Be sure to hold the display with the D-

Sub connector “in-line” with the D-Sub connector on the pressure transducer. 
The display should fully engage with the connector on the transduce. Insert 
the two screws supplied.

4.   Snugly tighten the two screws on the transducer display

3.3 ELECTRICAL CONNECTION

SEE CAUTION

1.   Connect the input/output connection using the D-Sub connector. (Connection 

pinouts are presented on Page 2 of this manual.

2.   Power may now be applied to the display unit.

3.4 DISPLAY ALIGNMENT

Orient the display by rotating the bezel. The bezel can be rotated 290°.

3.5 MODE SETTIING

The display’s mode of operation can be switched from, or to, either PSIA or KPa.

Full scale pressure range is factory set. Refer to the part number for the range of 
the unit installed.

3.6 CALIBRATION

Calibration is not required for the SLR Pressure Transducer Display. The pres-
sure transducer potentiometers should be used to adjust the zero of the pres-
sure transducer.

3.7 REMOVAL AND REPLACEMENT

1.   Remove power to the display.

2.   Disconnect the Input/Output connection.

3.   Remove the two retaining screws located on the display unit. Keep the screws 

with the unit.

4.   Pull the display unit upwards to release it from the pressure transducer.

5.   If replacing the display, follow Mechanical Installation steps.

CAUTION

!

!

Failure to follow these proce-
dures may adversely affect 
the product’s performance 
and could void the product 
warranty. Inspect but DO NOT 
unwrap any parts until instal-
lation. Contact your Celerity 
representative with any prob-
lems.

CAUTION

!

!

To prevent damage to the trans-
ducer or the display, do not 
over tighten the screws.

CAUTION

!

!

Dispose of a used display unit 
in compliance with local regu-
lations.

3.0

_INSTALLATION

Summary of Contents for Surface Line Rotatable Series

Page 1: ...Surface Line Rotatable SLR Series Pressure Transducer Display User Guide Celerity Inc 915 Enterprise Boulevard Allen TX 75013 USA T 1 972 359 4000 F 1 972 359 4100 A331289 REV 001 8 07 ...

Page 2: ... _2 1 OPERATING SPECIFICATIONS 2 _3 0 INSTALLATION 3 _3 1 SLR PRESSURE TRANDUCER DISPLAY DIMENSIONAL DRAWING PINOUTS 3 _3 2 MECHANICAL INSTALLATION 4 _3 3 ELECTRICAL CONNECTION 4 _3 4 DISPLAY ALIGNMENT 4 _3 5 MODE SETTING 4 _3 6 CALIBRATION 4 _3 7 REMOVAL AND REPLACEMENT 4 _4 0 WARRANTY 5 _CONTENTS ...

Page 3: ...rface substrate It is designed to interface via 15 Pin D connector to SS2 transducers with 30 and 100 PSI full scale pressure ranges The SLR utilizes a 3 5 segment light emitting diode LED with two selectable engineering units PSIA and KPa and can be rotated to be positioned at all four quadrants The SLR is electrically interfaced using a standard high density 15 pin D Sub HD15P connector The devi...

Page 4: ...ay Count Thermal Sensitivity 0 2 Display Counts C Operating Conditions Range 30 PSI 207 KPa 100 PSI 689 KPa Temperature Range 30 to 140 F 0 to 60 C Visual Display 290 Total Rotation Electrical Power Requirements Current Powered 3 92 20 4 mADC Voltage Powered 13 0 to 32 VDC Output 3 1 2 Digit Red 7 Segment LED 0 30 7 6mm High Input Requirements 0 5 VDC 0 10 VDC 4 20mA Minimum 4 95 9 9 3 92mADC Typi...

Page 5: ...ed 2 Pressure Signal 2 Pressure Signal 7 Power Input 7 Power Input 5 121 Power Supply Common Signal Common 5 121 Power Supply Common Signal Common Current Mode Pinout DB15 MALE PIN NUMBER SIGNAL DB15 FEMALE PIN NUMBER SIGNAL 1 2 3 4 6 8 9 15 Not Used 1 2 3 4 6 8 9 15 Not Used 7 Pressure Signal 7 Pressure Signal 5 121 Signal Common 5 121 Signal Common 47 75 1 88 30 48 1 20 28 58 1 13 Ø 28 58 1 13 P...

Page 6: ...can be switched from or to either PSIA or KPa Full scale pressure range is factory set Refer to the part number for the range of the unit installed 3 6 CALIBRATION Calibration is not required for the SLR Pressure Transducer Display The pres sure transducer potentiometers should be used to adjust the zero of the pres sure transducer 3 7 REMOVAL AND REPLACEMENT 1 Remove power to the display 2 Discon...

Page 7: ...ance contact Celerity Technical Support at 972 359 4000 Celerity is a trademark of Celerity of Celerity Inc All other product or service names mentioned in this document may be trademarks of the companies with which they are associated System descriptions are typical and subject to change without notice 2007 Celerity Inc A331289 Rev 001 08 07 CELERITY INC 915 Enterprise Boulevard Allen TX 75013 US...

Reviews: