Cecomp
®
Test +
Battery Powered Digital Pressure Gauges, User Configurable
CTP3B
Ranges and Resolution
See table below for ranges and resolution
User configurable for listed engineering units
Resolution is fixed for each engineering unit
Accuracy
±0.1% full scale ±1 least significant digit
Accuracy includes linearity, hysteresis, repeatability
Sensor hysteresis: ±0.015% FS, included in accuracy
Sensor repeatability: ±0.01% FS, included in accuracy
Display
3 readings per second nominal display update rate
4 digit LCD, 0.5"H and 5 character 0.25"H alphanumeric
White LED backlight active for 1 minute with button keypress
Backlight user configurable for AUTO (1 minute), ON, or OFF
Batteries, Battery Life, Low Battery Indication
2 AA alkaline included
Approx. 150-1500 hours depending on backlight usage
Low battery symbol on display
Controls and Functions
Three button keypad: Zero/Clear/Up, On/Off, Memory/Down
Passcode protected calibration, engineering unit selection,
auto shutoff time, min/max setup
Min/Max Memory
Minimum and/or maximum readings stored in memory
Readings cleared or stored at shutoff
User configurable
Calibration
Zero button for gauge reference ranges
Non-interactive zero, span, and linearity, ±10% of range
Auto Shutoff
Factory default: 5 minutes
User selectable: 1 minute to 8 hours or manual on/off
Weight
Gauge:
9 ounces (approximately)
Shipping:
1 pound (approximately)
Housing Materials
Epoxy powder coated aluminum case with case stiffener,
ABS/ polycarbonate bezel, coated circuit boards for humidity
resistance, front and rear rubber gaskets, polycarbonate label
NEMA 2 (IP51)
Includes rubber boot
Connection and Material
1/4" NPT male fitting
All wetted parts are 316L stainless steel
Overpressure, Burst, Vacuum Service
3000 psig sensor: 5000 psig overpressure
5000 psig sensor: 7500 psig overpressure
All others:
2 X pressure range overpressure
Burst pressure:
4 X sensor pressure rating, or 10,000 psi,
whichever is less
Vacuum service:
15 psia, ±15 psig, 15 psig, 30 psia,
100 psig, 100 psia, 200 psig sensors
Environmental Temperatures
Storage temperature:
–40 to 203°F (–40 to 95°C)
Operating temperature:
–4 to 185°F (–20 to 85°C)
Sensor compensated range: 32 to 158°F (0 to 70°C)
Dimensions
3.67"W x 3.19"H x 2"D with boot, not including fitting
© 02-21
cecomp.com
1220 American Way Libertyville, IL 60048
800-942-0315
CECOMP
Div. of
A
A
BSOLUTE
BSOLUTE
P
P
ROCESS
ROCESS
I
I
NSTRUMENTS
NSTRUMENTS
cecomp.com/plus
H H H H H H
H H H H H
H H H H H H
H H H H H
H H H H H H
H H H H H
H H H H H H
H H H H H
H H H H H H
Made in USA
O
±0.1% Test Gauge Accuracy
O
Ruggedized Design
O
Red Rubber Boot Included
O
316L Stainless Steel Wetted Parts
O
Selectable Units and Auto Shutoff Times
O
White LED Display Backlight
O
Min/Max Memory
Calibration Documentation Options—add to end of model number
CD
Calibration data; 5 test points and date
NC
NIST traceability documentation, 5 points and date
Vacuum
Range/Resolution
Selectable Units
CTP3B15PSIVAC
0 to 14.70 psig vacuum
psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2 cmH2O, g/cm2, kg/cm2, atm, bar, mbar, MPa, kPa
Vacuum/Pressure
Range/Resolution
Selectable Units
CTP3B±15PSIG
–14.70 to 15.00 psig
psi, inHg/psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2, cmH2O, g/cm2, kg/cm2, atm, bar, mbar, MPa, kPa
CTP3B30V15PSIG
–29.92 inHg to 15.00 psig
psi, inHg/psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2, cmH2O, g/cm2, kg/cm2, atm, bar, mbar, MPa, kPa
CTP3B15V100PSIG
–14.7 to 100.0 psig
psi, inHg/psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2, cmH2O, g/cm2, kg/cm2, atm, bar, mbar, MPa, kPa
CTP3B30V100PSIG
–29.9 inHg to 100.0 psig
psi, inHg/psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2, cmH2O, g/cm2, kg/cm2, atm, bar, mbar, MPa, kPa
CTP3B15V200PSIG
–14.7 to 200.0 psig
psi, inHg/psi, inHg, torr, inH2O, ftH2O, oz/in2, kg/cm2, atm, bar, MPa, kPa
CTP3B30V200PSIG
–29.9 inHg to 200.0 psig
psi, inHg/psi, inHg, torr, inH2O, ftH2O, oz/in2, kg/cm2, atm, bar, MPa, kPa
Pressure
Range/Resolution
Selectable Units
CTP3B3PSIG
0 to 3.000 psig
psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2, mmH2O, cmH2O, g/cm2, kg/cm2, atm, bar, mbar, kPa
CTP3B5PSIG
0 to 5.000 psig
psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2, mmH2O, cmH2O, g/cm2, kg/cm2, atm, bar, mbar, kPa
CTP3B15PSIG
0 to 15.00 psig
psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2, cmH2O, g/cm2, kg/cm2, atm, bar, mbar, MPa, kPa
CTP3B30PSIG
0 to 30.00 psig
psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2, cmH2O, g/cm2, kg/cm2, atm, bar, mbar, MPa, kPa
CTP3B60PSIG
0 to 60.00 psig
psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2, cmH2O, g/cm2, kg/cm2, atm, bar, mbar, MPa, kPa
CTP3B100PSIG
0 to 100.0 psig
psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2, cmH2O, g/cm2, kg/cm2, atm, bar, mbar, MPa, kPa
CTP3B200PSIG
0 to 200.0 psig
psi, inHg, inH2O, ftH2O, oz/in2, kg/cm2, atm, bar, MPa, kPa
CTP3B300PSIG
0 to 300.0 psig
psi, inHg, ftH2O, oz/in2, kg/cm2, atm, bar, MPa, kPa
CTP3B500PSIG
0 to 500.0 psig
psi, inHg, ftH2O, kg/cm2, atm, bar, MPa, kPa
CTP3B1000PSIG
0 to 1000 psig
psi, inHg, ftH2O, kg/cm2, atm, bar, MPa, kPa
CTP3B2000PSIG
0 to 2000 psig
psi, inHg, ftH2O, kg/cm2, atm, bar, MPa
CTP3B3000PSIG
0 to 3000 psig
psi, inHg, ftH2O, kg/cm2, atm, bar, MPa
CTP3B5000PSIG
0 to 5000 psig
psi, kg/cm2, atm, bar, MPa
Absolute Reference
Range/Resolution
Selectable Units
CTP3B15PSIA
15.00 to 0 psi absolute
psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2, cmH2O, g/cm2, kg/cm2, atm, bar, mbar, MPa, kPa
CTP3B30PSIA
30.00 to 0 psi absolute
psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2, cmH2O, g/cm2, kg/cm2, atm, bar, mbar, MPa, kPa
CTP3B100PSIA
100.0 to 0 psi absolute
psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2, cmH2O, g/cm2, kg/cm2, atm, bar, mbar, MPa, kPa
CON14SS
Quick connector to install or remove gauge
without tools. 304 stainless steel, urethane
seal.
SCR14SS
Filter screen fitting keeps debris out of gauge
sensor. For food vacuum packaging applica-
tions. 303SS body, 100 micron 304SS screen.
Accessories—order separately
DPG-OK2, DPG-OK3, DPG-OK6
Pelican
®
brand high visibility orange heavy
duty waterproof cases. Models available for
storing 2, 3, or 6 gauges.