background image

©

 2014 CDI Meters, Inc. 

10.14 

moderate pressure, pressing squarely on the drill, and retracting the bit from time 
to time to clear the chips and cool the bit. After drilling the holes, remove any 
burrs that were formed and make sure the pipe is reasonably smooth in the area 
around the holes. Apply the “Holes in Pipe” decal so that it will be hidden when 
the meter is in place but will be revealed when it is removed. 

Installing the Meter 

First prepare the meter for the pipe orientation and the direction of flow. To 
switch the 5400 between horizontal and vertical orientation, loosen the screws on 
the back of the housing, rotate the display, then re-tighten the screws. Be careful 
to loosen the screws no more than necessary, and then tighten them securely. 
The flow arrow on the meter must be oriented to match the direction of flow in the 
pipe. If doing so will cause the display to be upside-down, remove the cover of 
the meter, lift out the display, rotate both 180° and re-install. 

Now mount the meter on the pipe and tighten the cap screws evenly. The torque 
should be sufficient to seal the gasket but not so great that it will crush the gasket 
or distort or over-stress a thin-walled pipe. 

Wiring the Meter 

Wiring must be in accordance with applicable codes and standards. In areas 
where electrical interference may be present, signal wiring should be shielded 
with the shield grounded remote from the meter. Signal cables must not be run in 
conduit or cable trays shared with power wiring. The main terminal block inside 
the meter is accessed by removing the cover and lifting out the display circuit 
board. 

P

OWER

 

Either use the dc wall-plug supply furnished with the meter, or connect the 
terminals marked “24V dc” to a distributed dc power supply and seal the unused 
power opening with the plastic plug provided. The meter will draw a maximum of 
250 mA. Unlike earlier 5200 and 5400 meters, the Rev. 3.0 meters do not 
connect the dc- terminal to the pipe on which they are mounted. Please note that 
18 Volt dc supplies furnished with some earlier CDI flowmeters do not provide 
the voltage required for Rev. 3.0 meters. 

R

EMOTE 

D

ISPLAY

 

If you are using the summing remote display (SRD), connect the three terminals 
marked “display” to the corresponding terminals in the remote display. The meter 
may be powered from the remote display if the cable is 22 gauge or heavier and 
the distance is no greater than 200 feet. 

M

ILLIAMP 

O

UTPUT

 

The 4 - 20 milliamp output is optically isolated within the meter and it is provided 
with a factory-installed jumper which allows it to source a milliamp signal 
powered from the meter. The resistance of the loop connected to the output 
should not exceed 600 Ohms.  

©

 2014 CDI Meters, Inc. 

10.14 

 

 

P

ULSE 

O

UTPUT

 

The pulse output is open-collector, that is, a transistor switch to the meter’s 
negative supply. To use it, connect it to the input of the receiving device; also 
connect that input through a 10K Ohm resistor to a suitable positive supply and 
connect the negative supply of the receiving device to the negative supply of the 
meter. If an isolated relay contact is required, install the CDI 5200-IPO isolated 
pulse output and wire it to the receiving device. 

S

ERIAL 

C

OMMUNICATION 

O

PTION

 

The RS-485 signal used with the Serial Communication option is wired to the 
special Serial Communication display board. The two threaded openings in the 
meter enclosure permit the signal to be wired from meter to meter without 
external splices. Please refer to the data sheet for the Serial Communication 
option. 

Reviews: