![ccei PF10J025 Technical Manual Download Page 7](http://html1.mh-extra.com/html/ccei/pf10j025/pf10j025_technical-manual_2575775007.webp)
PHENIX
Technical Manual
v2.0
CCEI
page 7
This screen is used to define the volume of the pH corrector container.
The volume of the container is adjustable from 0 to 25L. To define the volume, simply
press the
and choose the value using the
and keys.
To validate the set value, press the key. To exit the programme without validating
the mode, press the key.
NOTE: As PHENIX injects the product into the pool, the value previously defined is
decreased. When this value reaches 0, the PHENIX pump is blocked. To re-enable
operation of the pump, the user should reset the container volume as described below.
If the volume is set to 0, there will be no controls on the pH corrector container
This menu is used to calibrate the pH sensor. This operation must be carried out first
time the appliance is commissioned as stated in the Maintenance paragraph.
Calibration is carried out using the two solutions supplied with the PHENIX: a PH7
solution and a PH9 solution.
To begin calibration, press the key. The PHENIX then displays CAL. pH7:x.x?. This
means that the user should immerse the sensor in the PH7 solution, wait until the pH
value stabilises and then press the
key. If the value is correct, the calibration
procedure continues, if not, the Err. ! message flashes - this means you have to press
the
key to start the measurement again.
PHENIX then displays CAL. pH9:x.x?. This means that the user should immerse
the sensor (after having cleaned it) in the PH9 solution and then press the
key.
If the calibration is correct, the pH Calibration message displays. If not, the Err. !
message appears and the calibration should be repeated after having checked the
solutions and the condition of the sensor. If the problem persists, contact your
installation engineer.
Cont = 20.0L (pH)
pH Calibration