18
2. Attach a NEW handpiece and hold Tap-On
TM
Foot Pedal
for 2 seconds. If blinking stops, the system is ready
for use. Discard the old handpiece or return if within
warranty. If blinking remains, continue to the next
step.
3. Install and fully seat an insert into handpiece. Hold
Tap-On
TM
Foot Pedal for 2 seconds. If blinking stops,
the system is ready for use. If blinking remains,
continue to the next step.
4. Install and fully seat a NEW insert in handpiece and
hold Tap-On
TM
Foot Pedal for 2 seconds. If blinking
stops, system is ready for use. Discard old insert or
return if within warranty. If blinking remains, refer to
Section 11.2 Technical Support and Repairs to
have unit serviced as soon as possible.
• Slow Blinking (1 blink per second)
-- The system is not operating within factory
specifications.
1. Remove insert.
2. Turn Main Power Switch OFF, (O) position. Wait five
seconds. Turn unit ON, (I) position.
3. Operate Purge function.
4. If service indicator still blinks, refer to section 11.2
Technical Support and Repairs to have unit
serviced as soon as possible.
Symptom:
System operates: Service Indicator illuminated
1. Ensure that the base unit has adequate ventilation and
is not near a heat source (i.e. radiator, heat lamp,
sunlight or other heat producing operatory equipment).
2. Turn Main Power Switch to the OFF (O) position.
Allow system to cool for 10 minutes and turn system
to the ON (I) position. Verify light is not illuminated.
3. If light is still illuminated, refer to section 11.2
Technical Support and Repairs to have unit serviced as
soon as possible.
11 2 Technical Support and Repairs
For technical support and repair assistance call DENTSPLY
Professional Cavitron Care
SM
Factory Certified Service at 1-800-
989-8826 Monday through Friday, 8:00 A.M. to 5:00 P.M.
(Eastern Time). For areas outside the U.S., contact your local
DENTSPLY Professional representative.
SECTION 12:
warranty period
The Cavitron Plus Ultrasonic Scaler is warranted for TWO
YEARS from date of purchase. The Steri-Mate
®
Handpiece
enclosed with your system is warranted for SIX MONTHS from
date of purchase. Refer to the Warranty Statement Sheet
furnished with your system for full Warranty Statement and
Terms.
SECTION 13:
Specifications
Electrical Voltage Continuous (100-240 VAC)
Current
1.0 Amperes, Maximum
Phase
Single
Frequency
50/60 Hertz
Water Pressure 20 to 40 psig
(138 to 275 kPa)
Water Temperature < 25
o
C (77
o
F)
Air Pressure
65 to 100 psig
(448 to 600 kPa)
Water Flow Rate Minimum Setting (CCW) < 15 ml/min
Maximum Setting (CW) > 55 ml/min
Weight 4.4 lbs (2 Kg)
Dimensions
Height: 6 in. (15,24 cm)
Width: 9.5 in. (24,13 cm)
Depth: 8 in. (20,32 cm)
Handpiece Cable length: 6.5 ft. (2.0 M)
Auxillary Footswitch Cable length: 8 ft. (2.4 M)
Water Supply Line length: 8 ft. (2.4 M)
Footswitch
Protection Class IPX1. Not for operating
theatres
.
Remote
Communication Frequency: 2405 to 2480 MHz
Power: < 1mW
Channels: 16
Operating
Environment Temperature: 15 to
40 Deg.Celsius (59 to 104 Deg
Fahrenheit)
Relative Humidity: 30% to 75% (non-condensing)
Transport and Storage Conditions Temperature:
-40 to 70 Deg. Celsius (-40 to 158 Deg.
Fahrenheit)
Relative Humidity: 10% to 100% (non-
condensing)
Atmospheric Pressure: 500 to 1060 hPa