Page 10
Changing or Viewing Instrument Settings
Setup Screen
The setup screen tells you the following information: -
The detector the instrument is using
The parameter the instrument is measuring
The filter the instrument is using
The operating range of the instrument
Detector
The detector determines if the instrument is measuring the running rms
(Root Mean Square) value or the peak value.
If rms is selected then the display shows RMS.
If peak is selected then the disply shows PK.
Depending on the instrument type and the inetegration method selected, the
running rms is calculated as shown below: -
Running rms
Acceleration
(g)
GA2002
Arms
1
t
t
ta
( )
a
w
2
ta
( )
d
9.807
g
( )
t = instantaneous time (seconds)
= integration time of the measurent (seconds)
a
w
(ta) = instantaneous acceleration value
(ta) = time (seconds)
Running rms
Acceleration
(ms
-2
)
GA2003
Arms
1
t
t
ta
( )
a
w
2
ta
( )
d
ms
2
t = instantaneous time (seconds)
= integration time of the measurent (seconds)
a
w
(ta) = instantaneous acceleration value
(ta) = time (seconds)