background image

 
 

 
 

The STATUS button is used to display the status of the instrument. 

 

 
Press 

STATUS                                                                     BAT  80% 

 

Repeatedly pressing the 

STATUS

 button will scroll through the parameters 

listed 

below, depending on the instrument type. 

 

 

INSTRUMENT STATUS 

 All 

BAT.80% 

Percentage battery capacity remaining 

ALL 

dBA SLOW 

Indicates the current frequency and time weighting 

GA213/215 

VER2.0 

Indicates the firmware version number 

ALL 

RUNNING      
           

 

Indicates the instrument is running 

ALL 

PAUSED 

Indicates when the instrument is in pause mode (only 

displayed if instrument is in pause mode) 

ALL 

90/3 

Indicates the current criterion and exchange level 

GA256 

www.

castlegroup.co.

uk 

Scrolling web-site address 

ALL 

12 

 

The STATUS menu is therefore circular and can be easily scrolled through by 
repeated pressing of the 

STATUS 

button. Press the 

DISPLAY

 button at any time 

to return to normal operation. 

 

This format is repeated for the other functional buttons on your instrument.  

 

 

 

 

 

 

The calibrators recommended for use are the Castle GA601/607 calibrators, 

which supply 94dB/ 104dB (relative to 20µPa pressure) at a frequency of 1kHz. 

 
Turn the instrument on by pressing the 

POWER 

button. 

 
Press 

STATUS

 to check the battery condition.  Replace the battery if the 

indication is low.      

 

Press 

MENU                                               CAL    <OK 

 

 

Press 

OK 

and the instrument will now display the previous calibration level  

e.g. >94.0.  The calibration will be performed at the selected level  

e.g.94.0dB but may be changed in 0.1 dB steps to match the calibrator output 
by repeated pressing of the 

UP/DOWN 

arrow buttons or in 1dB steps by 

repeated pressing of the 

STATUS/ PLAY 

keys.  Ensure the calibrator is 

attached firmly to the microphone.  Ensure that the calibrator is set to the 

chosen level. 

 
Press 

OK

 and the instrument will calibrate to the level of the calibrator.  The 

display will show 

SETTLING

 while calibrating to the chosen level. When calibrated 

the display will show 

COMPLETE. 

Press 

OK

 and the display will return to LP for 

the GA213/215 and DOSE for the GA256. 

 

If the input level is not /- 3dB of the chosen reference level then the 

display will show CAL.FAIL.  Check to make sure the calibrator is switched on and 

emitting the correct level before proceeding as indicated above.   

 

The instrument is now ready to take measurements. 

 

CALIBRATION 

All 

Summary of Contents for GA213

Page 1: ...Notes ...

Page 2: ... Service Castle Group Ltd design and manufacture precision instruments which if treated with reasonable care and attention should provide many years of trouble free service In the event of a fault occurring during the warranty period the instrument should be returned to Castle Group Ltd in its original packaging or to an authorised agent Please enclose a clear description of the fault or symptom D...

Page 3: ...r the operation of the unit are as follows Temperature 10 C to 50 C Relative Humidity 30 to 90 Protect the unit from dust water extremes of temperature and humidity direct sunlight and air with a high salt or sulphur content Always turn the unit off after use Remove the battery from the instrument if it is not to be used for an extended period several days Do not use any solvents or cleaning agent...

Page 4: ...r GA601 GA607 and microphone Corrected level Actual level 0 05 t 20 Where t is the temperature in C Effects of Humidity Less than 0 5 dB over the range 25 to 90 relative humidity provided there is no Condensation relative to the value at 50 relative humidity and 40 C Storage range 0 to 90 relative humidity in the absence of condensation Vibration Effect From 20Hz to 1kHz at 1ms 2 no noticeable eff...

Page 5: ...essure 94 0 dB at 1kHz nominal reference level 55 120dB reference range Reference direction of incidence for all microphones perpendicular to the front face diaphragm surface of the microphone Detector Characteristics RMS and Peak Warm up time 10 minutes Operating range 10 C to 50 C Warning DO NOT subject the instrument to temperatures greater than 70 C or less than 20 C for any length of time 16 ...

Page 6: ...s press that button E g OK means press the OK button Since this manual covers the operating instructions for three instruments the shaded areas of the title serves to inform the user which functions are present on your instrument E g The following title informs you that this function is only applicable to the GA213 215 2 CHANGING RANGE GA213 215 Linear Operating Range GA213 GA215 65dB GA256 67dB R...

Page 7: ...ment for use a heavy duty alkaline battery type 6LR61 should be fitted such as a Procell Duracell MN1604 Ever Ready 6LF22 Gold Seal or equivalent The battery door is located on the bottom left hand side of the instrument Open the battery door cover by sliding the cover downward towards the bottom of the instrument The cover will now swing open exposing the battery compartment Insert the battery ob...

Page 8: ...configurations Press STATUS BAT 80 Battery status Press STATUS dBA SLOW Frequency Time weighting selected Press STATUS VER 1 0 Firmware version number Press STATUS WWW Cast website scrolling Press STATUS RUNNING SWITCHING ON AND OFF All 4 ACTIVATING BUTTONS All 13 The individual characteristic applies to each instrument unless specifically worded otherwise Instrument Standards GA213 IEC 61672 1 20...

Page 9: ...04dB relative to 20µPa pressure at a frequency of 1kHz Turn the instrument on by pressing the POWER button Press STATUS to check the battery condition Replace the battery if the indication is low Press MENU CAL OK Press OK and the instrument will now display the previous calibration level e g 94 0 The calibration will be performed at the selected level e g 94 0dB but may be changed in 0 1 dB steps...

Page 10: ...PLAY button The table on the following page is a breakdown of the display parameters and is accompanied by an explanation of each one DISPLAYING PARAMETERS All GA213 GA215 GA256 Lp SPL Lp SPL Lex Lmx Leq Plx Pmx Elapsed Time DO Elapsed Time Lmx Pr Overload if triggered Pmx Mr Lex Elapsed Time Le Pmx Overload if triggered Psq Overload if triggered 6 The Criterion and Exchange level can be changed b...

Page 11: ...ailable weighting choices A SLOW A FAST C SLOW or C FAST Note The symbol indicates the current selection as you move around the weighting options CHANGING RANGE GA213 215 10 FREQUENCY TIME WEIGHTING GA213 215 DISPLAY DESCRIPTION INSTRUMENT 93 7dBC Level Pressure Lp GA213 215 LEQ 93 7 Displays the LEQ GA215 00 01 33 Run time duration All Lmx 97 3 Maximum Lp GA215 Pmx 98 8 Maximum Peak GA213 256 DO ...

Page 12: ... button to return to LP mode Press PAUSE PLAY once PAUSED All functions can be held so that they do not update by pressing the PAUSE button Once PAUSED the display may be read as normal allowing the user to take note of any required readings For normal operation of the instrument to continue press the PAUSE PLAY button once more and the instrument will briefly show PAUSED before returning to the c...

Reviews: