background image

3 – ANhyDRoUS CoNfIGURATIoN – continued

3 – ANhyDRoUS CoNfIGURATIoN – continued

3.4  Nh

3

 Drive Set-up

a)  Select NH3 Drive [Yes]
b)  Press ‘Setup’
c)  Assign NH3 Drive Serial 

Numbers (starting w/
modules on LH side)

d)  Select Drive Type [Servo 

or PWM]

e)  Select Master Valve Type
f)  Select Pump Disarm
fg) Select Sec Off Behavior  

[Lock at Last or Turn Off]

h)  Enter Drive Meter Cal 

Number  
(Pulses/Gal)

Note:

 Pulses/gallon will be found on the liquid flow meter. Some 

flow meters are measure in pulses/10 gallons. If this is the case, 
divide that value by 10 to find pulses/gallon.

i)  Press ‘Done’

3.5  Master Switch Box  
(If Equipped with External Switch Box)

a)  Select Master Sw Box [Yes]
b)  Press ‘Setup’
c)  Verify Serial Number of 

Switch Box

d)  Select Footswitch (if 

installed)

e)  Press “Done’

3.6  Implement Switch  
(If equipped with Toolbar Mounted Implement Switch)

a)  Select Imp Switch [Yes]
b)  Press ‘Setup’
c)  Select Imp Switch  

Serial Number

d)  Select Switch Polarity 

Note: 

Determine by raising and lowering the implement and 

watch the Implement Status Arrow in Status/Warning Area for 
proper operation.

e)  Press ‘Done’

Note:

 #1 EHR work switch on Magnum & Steiger tractors can be 

used as an implement switch (the #1 EHR must be cycled once 
after each start up to display the status arrow).

3.7  Section Switch Box (If equipped with  
External Section Switch Box or Desire  
Manual valve Section Control through Run Screens) 

a)  Select Section Switch Box 

[Yes]

b)  Press ‘Setup’
c)  Select Config Mode [Auto] 

or [Manual] for custom

d)  Verify Sw Box Serial 

Number (if equipped w/ 
external switchbox)

Note:

 If no external switchbox 

is installed, User Defined  
Windows 

can be assigned  

to a Run Screen  
(

Toolbox>layout

)

4 – WoRk CoNDITIoN – REQUIRED foR oPERATIoN

4 – WoRk CoNDITIoN – REQUIRED foR oPERATIoN 

(continued)

Work Condition>layer

IMPoRTANT!

 The settings below are linked to a 

work condition. These must be selected or checked 

whenever a work condition is created/changed: product type, 
application rate, drive settings, product delay, product layer 
control and product control.

NoTE!

 A work condition name could be for a crop type, field 

condition, or weather condition, etc.

NoTE!

 This setup is required for logging/ mapping data and 

using Overlap &  Boundary Control.

4.1 Preparation

a)  Insert a data card in the display
b)  Create/select a Grower/Farm/Field/Task and Crop Type 

(

Performance > Profile

)

4.2 Product Setup - continued

NoTE!

 App Rate will not  

be able to be adjusted 
outside of this range

NoTE!

 App Rates are 

measured in pounds of 
actual Nitrogen, NOT 
pounds of NH

3

.

4.2 Product Setup

Toolbox> Product

a)  Name the product (ex. NH3)
b)  Select form type for product [Granular]
c)   Select Usage [Fertilizer]
d)  Enter Default Application Rate
e)   Enter Minimum Application Rate & Maximum  

Application Rate

4.3 Product layer Assignment

Work Condition> layer

Assign a product to a control section  
of the Applicator.

a)  Select/Create a  

Work Condition

b)  Select Layer 1 Control 

Type [AccuControl 
Anhydrous]

c)  Select Product for  

Layer 1 Control

d)  Assign Additional Layers  

as needed

4.4  Controller Setup – Anhydrous (if equipped)

Work Condition>Control

 

a)  Verify Implement
b)  Verify Work Condition
c)  Select Controller 

[Anhydrous]

d)  Product Delay – Default  

1.0 sec. (see section 7  
for Product Delay 
Measurement procedure)

e)  Enter the Minimum 

Speed (if ground speed 
drops below this speed, 
applicator will apply at 
this set speed)

 f)  Enter value for Off-target  

Alarm Limit 

Reviews: